3
LDS-750-01 4107 9513 61.00 Page 1 of 3 11 January 2019 CLAW SIMPLEX 2.3 - 5.4 BASIC HORIZONTAL TANK MOUNT BeaconMedæs • 1059 Paragon Way, Rock Hill, SC 29730 • Phone: (803) 817-5600 • Fax: (803) 817-5750 • www.beaconmedaes.com VerusLab “Oil-Less” Claw Laboratory Vacuum Simplex Horizontal Tank Mount Basic (PLC) Control System 2.3 - 5.4 HP SPECIFICATION Vacuum System The Laboratory Oil-Less Claw vacuum package features a com- mon base with single point connections for the electrical panel, intake, and discharge. The common base is sized to fit through standard 34.5” doorway. Designed and manufactured with ISO 13485 processes, each system is completely tested before shipment and includes: • One “oil-less” claw rotary vacuum pump with one motor • Integral pre-wired control panel • Horizontal air receiver with full-size three-valve bypass, sized for appropriate demand Vacuum Pump Each pump is a direct driven, non-contacting claw type, capable of operating continuous duty at 28.4” Hg at sea level. The pumping chamber is oil free. The pump is completely air-cooled with no water requirements. Each pump contains: 5 micron inlet air filter • Vacuum relief valve Check valve to prevent backflow through off-cycle units • Flexible connector and isolation valve • High discharge temperature sensor • Oil drain valve and oil sight glass • Each pump is mounted on vibration isolators Vacuum Pump Motor Motors are continuous duty, C-face, TEFC, 3450 RPM, available in 60 hertz (208V, 230V, or 460V) and 50 hertz (380V), 3-phase electrical service. Automatic Purge System The vacuum pump is equipped with an automatic purge system to flush any gases from the pump to prevent condensation as the pump cools. The purge system incorporates a 24V electric controlled automatic isolation valve and controls. The pump purge time is field adjustable from 5 to 15 minutes with factory setting at 7 minutes. This vacuum pump system requires no air supply. Vacuum Receiver The vacuum receiver is ASME Code stamped, and rated for a minimum 200 PSIG design pressure. The receiver has a full-size three-valve bypass system to allow for draining of the receiver without interrupting the vacuum service. A manual drain is provided on the receiver. Intake Piping Each vacuum pump has a factory piped intake with integral flex connector, isolation valve with 24V electric actuator, and check valve. Interconnecting piping consists of powder-coated steel tubing and flanges. Basic Control System The basic control system is U.L. labeled. Additional components include: • NEMA 12 control panel enclosure • Circuit breaker disconnect for the motor with external operator • Full voltage motor starter with overload protection • 24V control circuit • Visual alarms for high discharge temperature shutdown with isolated contacts for remote alarm. Hand-Off-Auto lighted selected switch • Runtime hourmeter • Vacuum gauge Installation The installation of this vacuum technology is limited to 6,000 feet maximum elevation above sea level. For installation of this equipment above this elevation, please contact the factory.
Transcript
Page 1: VerusLab “Oil-Less” Claw Laboratory Vacuum Simplex Horizontal … · 2020. 2. 26. · Capacity2 @ 19” HgV (scfm/ system) Capacity2 @ 25” HgV (scfm/ system) End Pressure inch

LDS-750-014107 9513 61.00

Page 1 of 3 11 January 2019

CLA

W S

IMPL

EX 2

.3 -

5.4

BASI

CH

ORI

ZON

TAL

TAN

K M

OU

NT

BeaconMedæs • 1059 Paragon Way, Rock Hill, SC 29730 • Phone: (803) 817-5600 • Fax: (803) 817-5750 • www.beaconmedaes.com

VerusLab “Oil-Less” Claw Laboratory VacuumSimplex Horizontal Tank Mount

Basic (PLC) Control System2.3 - 5.4 HP

SPECIFICATION

Vacuum SystemThe Laboratory Oil-Less Claw vacuum package features a com-mon base with single point connections for the electrical panel, intake, and discharge. The common base is sized to fit through standard 34.5” doorway. Designed and manufactured with ISO 13485 processes, each system is completely tested before shipment and includes:

• One “oil-less” claw rotary vacuum pump with one motor• Integral pre-wired control panel• Horizontal air receiver with full-size three-valve bypass, sized

for appropriate demand

Vacuum PumpEach pump is a direct driven, non-contacting claw type, capable of operating continuous duty at 28.4” Hg at sea level. The pumping chamber is oil free. The pump is completely air-cooled with no water requirements. Each pump contains:

• 5 micron inlet air filter• Vacuum relief valve• Check valve to prevent backflow through off-cycle units • Flexible connector and isolation valve• High discharge temperature sensor• Oil drain valve and oil sight glass• Each pump is mounted on vibration isolators

Vacuum Pump MotorMotors are continuous duty, C-face, TEFC, 3450 RPM, available in 60 hertz (208V, 230V, or 460V) and 50 hertz (380V), 3-phase electrical service.

Automatic Purge SystemThe vacuum pump is equipped with an automatic purge system to flush any gases from the pump to prevent condensation as the pump cools. The purge system incorporates a 24V electric controlled automatic isolation valve and controls. The pump purge time is field adjustable from 5 to 15 minutes with factory setting at 7 minutes. This vacuum pump system requires no air supply.

Vacuum ReceiverThe vacuum receiver is ASME Code stamped, and rated for a minimum 200 PSIG design pressure. The receiver has a full-size three-valve bypass system to allow for draining of the receiver without interrupting the vacuum service. A manual drain is provided on the receiver.

Intake PipingEach vacuum pump has a factory piped intake with integral flex connector, isolation valve with 24V electric actuator, and check valve. Interconnecting piping consists of powder-coated steel tubing and flanges.

Basic Control SystemThe basic control system is U.L. labeled.

Additional components include:• NEMA 12 control panel enclosure• Circuit breaker disconnect for the motor with external

operator• Full voltage motor starter with overload protection• 24V control circuit• Visual alarms for high discharge temperature shutdown with

isolated contacts for remote alarm.• Hand-Off-Auto lighted selected switch• Runtime hourmeter• Vacuum gauge

InstallationThe installation of this vacuum technology is limited to 6,000 feet maximum elevation above sea level. For installation of this equipment above this elevation, please contact the factory.

Page 2: VerusLab “Oil-Less” Claw Laboratory Vacuum Simplex Horizontal … · 2020. 2. 26. · Capacity2 @ 19” HgV (scfm/ system) Capacity2 @ 25” HgV (scfm/ system) End Pressure inch

2

4

5

6

1 Full-size 3-valve receiver bypass.

2 Oil sight glass and filler plug with ease of access for annual oil changes.

3 Isolation valve with 24 V electric actuator.

4 Flex connectors on vacuum inlet to isolate pump vibration from facility pipeline.

5 120-gallon horizontal receiver.

6 Individual 5 micron air inlet filter per vacuum pump to protect pump from incoming particulates.

7 Automatic Purge System.

8 Through the door disconnects for pumps for safety during service.

9 Single point vacuum inlet connection.

9

1

87

3

LDS-750-014107 9513 61.00

Page 2 of 3 11 January 2019

Standard Configuration2.3 - 5.4 HP Simplex Claw

Page 3: VerusLab “Oil-Less” Claw Laboratory Vacuum Simplex Horizontal … · 2020. 2. 26. · Capacity2 @ 19” HgV (scfm/ system) Capacity2 @ 25” HgV (scfm/ system) End Pressure inch

1

1

2

2

3

3

4

4

5

5

6

6

7

7

8

8

A A

B B

C C

D D

.XX .02

.XXX .005.XXXX .0005ANGLES .5

1 of 2

1:9

DSheet

Date:Revised: Date: Drawn:

ARE IN INCHES AND TOLERANCES ARE:UNLESS OTHERWISE SPECIFIED ALL DIMENSIONS

DOCUMENTDO NOT SCALE THIS

Scale:

Description:

GENERAL ASSEMBLY SX 5.4 HPLAB CLAW VAC HTM 120G BASIC

BEACONMEDÆS

4107 8567 70Drawing Number:

Rev:DN:HOP180918

TAR 11/26/2018

N/A

INSPECTION CHECKLISTUNLESS OTHERWISE SPECIFIED:

ALL ITEMS MARKED BY ONE OF THE ABOVE SYMBOLS ARE TO BE INSPECTED EQUIPMENT TO BE USED DURING RECEIVING INSPECTION SHALL BE PER QUALITY SYSTEM PROCEDURES

is loaned without the expressed written permission from BeaconMedæs Engineering.BeaconMedæs and may not be used for other than the purpose for which itThis drawing and the information contained thereon remain the property of

Form F-007 Rev. 01

TANK DRAIN 1/2" FNPT 67.75

54.38

30.12INLET

1-1/2" FNPT

1" FNPT DISCHARGE CONN(FLEX HOSE SHIPPED LOOSE)

30.12

54.38

67.75

1” FNPT DISCHARGE CONN (FLEX HOSE SHIPPED LOOSE

INLET 1-1/2” FNPT

TANK DRAIN 1/2” FNPT

BeaconMedæs 1059 Paragon Way Rock Hill, SC 29730 Phone: (803) 817-5600 Fax: (803) 817-5750 www.beaconmedaes.com

LDS-750-014107 9513 61.00

Page 3 of 3 11 January 2019

Standard Configuration2.3 - 5.4 HP Simplex Claw

Notes: • Discharge muffler ships loose• Allow 36 inches in front of control panel for maintenance and ventilation, all other sides require 24 inches of clearance.• Additional drawings/diagrams available for download at www.beaconmedaes.com.

Notes: 1. Normal operating conditions at a maximum ambient of (41°C) 105° F. Consult factory for higher ambient conditions.2. Capacity measured at reference conditions of absolute inlet pressure 1 bar (14.5 psi), intake air temperature 20°C (68° F).

System capacities include one pump running.

Vacuum System Specifications1 @ Sea Level

System Model No. Hz HP

Nominal Pumping

Speed(acfm/pump)

Capacity2 @19” HgV(scfm/

system)

Capacity2 @25” HgV(scfm/

system)

End Pressure

inch of HgV(Torr)

SystemBTU/Hr

db(A) per

Pump

SystemWeight

(lb)

System FLA

208V 230V 380V 460V

LVHT02S-120H-BS 50 2.3 37 13 6 28.4 (45) 4,072 70 1,043 N/A N/A 3.3 N/A

LVHT02S-120H-BS 60 2.3 44 16 7 28.4 (45) 4,072 70 1,043 7.3 6.6 N/A 3.4

LVHT03S-120H-BS 50 3.2 48 18 8 28.4 (45) 6,108 70 1,051 N/A N/A 4.3 N/A

LVHT03S-120H-BS 60 3.2 57 21 9 28.4 (45) 6,108 70 1,051 10.1 9.0 N/A 4.6

LVHT04S-120H-BS 50 4.6 66 24 11 28.4 (45) 8,144 79 1,011 N/A N/A 5.5 N/A

LVHT04S-120H-BS 60 4.6 79 29 13 28.4 (45) 8,144 79 1,011 10.7 9.8 N/A 5.0

LVHT05S-120H-BS 50 5.4 86 32 14 28.4 (45) 10,944 79 1,052 N/A N/A 6.1 N/A

LVHT05S-120H-BS 60 5.4 103 38 17 28.4 (45) 10,944 79 1,052 14.4 13.2 N/A 6.7


Recommended