RESULTS AND DISCUSSION
4.1 Screening of Urease producing Microbes
Bacillus badius MTCC 8082, Bacillus sphaericus MTCC 5100 and Cornybacterium
sp MTCC 8143 were screened for maximum urease activity by Nesslerization method
(Kayastha et al., 1995). The urease enzyme was found to be intracellular in all the
three isolates and the maximum enzyme activity found at respective incubation period
from 200 ml broth is given in table 4.1.
Table 4.1: Comparison of enzyme activity (IU) of isolates
Micro-Organism Enzyme Activity
(IU)
Incubation
Period (hrs)
Bacillus badius MTCC 8082 13.80 ± 1.120 21
Bacillus sphaericus MTCC 5100 53. 87 ± 1.1.41 24
Cornybacterium sp MTCC 8143 0.53 ± 0.932 18
The enzyme activity was calculated taking NH4+ as the standard curve as shown in
figure 4.1
RESULTS AND DISCUSSION
86
Figure 4.1: Standard curve of NH4+ ions (μg ml
-1)
4.2 Kinetic Characterization
As the enzyme activity of Cornybacterium sp MTCC 8143 was found to be very low
as compared to the other two isolates the further work was proceeded with Bacillus
badius MTCC 8082 and Bacillus sphaericus MTCC 5100. The Km and Vmax value of
both the isolates were evaluated in free and immobilized state in the absence and
presence of Pb2+
ions and was found as given in table 4.2 and 4.3.
Table 4.2: Kinetic parameters for Bacillus badius
Parameters Free Enzyme Immobilized Enzyme
Without Pb2+
With 96.6
nanomoles
Pb2+
Without Pb2+
With 96.6
nanomoles
Pb2+
Km (mM) 7.69 ± 0.6 7.69 ± 0.4 5.0 ± 0.7 5.0 ± 1.0
Vmax (µM/min) 110 ± 3.2 76.9 ± 2.9 75.0 ± 3.4 20 ± 3.5
RESULTS AND DISCUSSION
87
As per the change in Km and Vmax values in the absence and presence of lead it was
observed that the lead is a non- competitive inhibitor of urease as it does not cause
any change in Km but reduces Vmax drastically to half the value. Also the
immobilization by the hydrosol- gel method decreased the Km indicating a higher
enzyme – substrate affinity after the immobilization. As the studies with lead in the
assay reaction showed non- competitive inhibition of urease, a constant concentration
of 100mM urea was selected for biosensor development. The urease containing
membranes were stored at 4ºC and a decrease of 14.29 % enzyme activity was
observed after 2 months.
Table 4.3: Kinetic parameters for Bacillus sphaericus
Parameters Free Enzyme Immobilized Enzyme
Without Pb2+
With 4.8 X
10-2
nanomoles
nM Pb2+
Without Pb2+
With 4.8 X
10-2
nanomoles
Pb2+
Km (mM) 8.12 ± 0.112 5.83 ± 0.163 5.1 ± 0.141 2.90 ± 0.431
Vmax (µM/min) 139.64 ±8.980 96.45 ±7.850 55.0 ± 7.071 25.6 ± 6.222
As observed the immobilization of the enzyme in hydrosol - gel matrix reduced the
Km from 8.12 ± 0.112 mM to 5.1 ± 0.141 mM suggesting that this immobilization
matrix is favorable for urease immobilization and may be positioning the enzyme’s
active site at a more substrate accessible orientation. The lead was found to be
uncompetitive inhibitor of urease in case of Bacillus sphaericus as it decreased the Km
RESULTS AND DISCUSSION
88
and Vmax of the enzyme with the same factor (Loeppky, 2001). The Bacillus
sphaericus is a novel isolate with better kinetic properties than jack bean and Bacillus
pasteurii (Verma and Singh, 2003), hence gives lower limit of detection for heavy
metals (Verma and Singh, 2006). The uncompetitive inhibition of urease by lead
suggests that unlike Cu2+
and Ag2+
which bind to nitrogen of histidine and possibly
oxygen of aspartic and glutamic acid residues in the active site of urease (Saboury et
al., 2010), lead binds to thiol group of cysteine residues present in the flap which
causes modification in cysteine residues due to which the flap loses its mobility and
activity of urease is inhibited (Krajewska and Zaborska, 2007).
4.3 Lead Inhibition Studies
The inhibition effect of lead on urease activity was studied by incubating the enzyme
with urea along with different concentrations of lead in standard Nesslerization
method. It was observed that the Baciilus badius urease was inhibited by lead at the
lowest concentration of 0.22 nanomoles (1.2 µM) in free system (table 4.4) and the
linear range of inhibition was from 0.22 to 22 nanomoles (1.2 – 120 µM) in a
logarithmic pattern (figure 4.2).
Table 4.4: Inhibition affect of lead on Bacillus badius urease
Lead Conc.
(Nanomoles )
% Inhibition
0.22 27.63 ± 2.78
0.87 38.00 ± 1.93
2.2 42.11 ± 1.96
RESULTS AND DISCUSSION
89
8.7 55.24 ± 2.19
22 66.67 ± 2.12
Figure 4.2: Linear range of inhibition of lead on Bacillus badius urease
Since the lowest inhibition against lead (using Bacillus badius urease) was found to
be 0.22 nanomoles (1.2 µM), the isolate was further used to develop simple
colorimetric and ISE based potentiometric biosensor for lead in a hope to lower down
the detection limit of lead in milk to permissible levels (96.6 nM or 20 ng ml-1
) using
sensitive detection systems.
The inhibition studies were also conducted on Bacillus sphaericus using the same
Nesslerization method. The inhibition was observed upto 0.87 picomoles (4.83 nM) of
lead concentration with this isolate which was a very promising signal for developing
lead biosensor. The linear range of inhibition was observed from 0.87 to 87 picomoles
RESULTS AND DISCUSSION
90
(4.83 – 483 nM) as shown in figure 4.3. So much sensitivity towards lead could be
attributed to its better kinetic properties as compared to other microbial urease (Verma
and Singh, 2003) and the efficiency of the microbe to detect heavy metals like Cu2+
and Ni2+
to nanomolar levels as observed by Verma and Singh (2006).
Figure 4.3: Linear range of inhibition of lead on Bacillus sphaericus urease
The inhibition affect of other heavy metals was also studied on both the isolates and it
was observed that lead has low inhibition efficiency as compared to other highly toxic
metals like Hg2+
and Cu2+
. The pattern of heavy metal inhibition found in B.badius
and B. sphaericus was Hg2+
˃ Cd2+
˃ Ni2+
˃ Cu2+
˃ Zn2+
˃ Co2+
˃ Fe2+
˃ Pb2+
and Hg2+
˃ Cu2+
˃ Cd2+
˃ Zn2+
˃ Co2+
˃ Fe2+
~ Pb2+
respectively. This inhibition
pattern shows the need of a step to nullify the affect of all other heavy metals on
urease.
4.4 Development of Colorimetric Biosensor for lead
The urease enzyme (Bacillus badius) was immobilized by hydrosol - gel method on
nylon membrane and a simple colorimetric method was devised for lead detection.
RESULTS AND DISCUSSION
91
Phenol red was used as the pH indicator, and the change in color after 10 mins
incubation of the membrane in lead containing urea solution was observed. There was
a clear distinction in the color of the sample before and after the reaction as shown in
figure 4.4 and the difference in color production with different lead concentrations
was detected spectrophotometrically by taking phenol red absorbance at 508 nm.
A B
Figure 4.4: Color of the reaction mixture before and after the reaction. A- Before
reaction, B- After reaction
In synthetic system with immobilized enzyme linear range detection from 48 – 96.6
nanomoles (48.3 – 96.6 µM) was observed (figure 4.5) and the lower limit of
detection was found to be 38.6 nanomoles (38.6 µM).
RESULTS AND DISCUSSION
92
Figure 4.5: Linear range of detection of lead in synthetic system
The same system was applied for the dection of lead in acid exracted spiked milk
samples. The linear range of detection was found to be 38.6 to 96.6 nanomoles (38.6 –
96.6 µM) as shown in figure 4.6 and lower limit of detection achieved in milk
samples was 38.6 nanomoles (38.6 µM).
Figure 4.6: Linear range of detection of lead in spiked milk
RESULTS AND DISCUSSION
93
The reliability of the sensor was checked by recovery studies on spiked milk samples
as shown in Table 4.5. It was observed that the lead found to the sample was close to
the added amount as detected by the developed system.
Table 4.5: Percentage recovery in spiked milk samples
Milk Samples Lead Added
(nanomoles)
Lead Found
(nanomoles)
% Recovery
Milk Sample 1 48 58 120.83
Milk Sample 2 68 77 113.23
Milk Sample 3 87 97 111.49
The present biosensor is based on easily extractable, cost effective urease enzyme
having lower limit of detection for lead as compared to the expensive enzymes used
by other workers. Fennouh et al. (1998) developed an amperometric biosensor with
L- lactate dehydrogenase (LDH) and L - lactate oxidase. The biosensor had a Pb2+
detection limit of 0.2 µM and the enzyme membranes were stable for more than 2
months at 4ºC. Kremleva et al. (1999) developed an electrochemical method for
determining Pb2+
using an amperometric cysteine desulfhydrolase tissue biosensor,
detection limit for lead was 20 nM for this system. The present developed biosensor
has a lower detection limit than the optical sensors developed by Kuswandi (2003)
and Tsai et al. (2003) with a detection limit of 10 µM and 100 µM respectively. The
present biosensor, developed by the authors has an application in milk samples as
worked by Babkina and Ulakhovich (2004) using ssDNA (detection limit 0.1nM).
Although the detection limit of our biosensor is higher than the above one, it has an
RESULTS AND DISCUSSION
94
advantage of low cost and easy pretreatment of the sample. The present biosensor
has far more better detection limit than the one developed by Rodriguez et al.
(2004a) which was based on urease-glutamic dehydrogenase amperometric assay for
heavy metal screening in polluted samples. The biosensor detected Pb level more
than 2.4 X 10-1
mM. Ogonczyk et al. (2005) developed a screen printed disposable
urease based biosensor for detection of heavy metals and reported a detection limit of
1 mM of lead in 20 min response time. Haron and Ray (2006) developed an optical
biosensor with very low detection limit of 4.83 nM, but have been applied on
drinking water samples only. The detection range obtained by the developed
biosensor is economical and comparable to Bagal – Kestwal et al. (2008). They
obtained a detection limit of 30 nM in 10 min contact time by studying the inhibition
of invertase and glucose oxidase on electrochemical transducer. Recently DNAzyme
based sensors has attained a detection limits ranging from 10 nM to 500 nM (Liu and
Lu, 2000; Wei et al., 2008; Wang and Irudayaraj, 2011). The low detection limit of
the present biosensor could be attributed to the high sensitivity of the urease form
Bacillus badius (Verma et al., 2010) towards the heavy metal ions.
A very simple enzyme inhibition based biosensor has been constructed for the
estimation of lead in milk samples after a brief treatment of milk samples. The
developed biosensor detects lead specifically in the presence of other metal cations in
milk sample, after preconcentration with lead specific column. The present biosensor
could be used for the estimation of lead in alloys.
4.5 Development of Ion Selective Electrode (ISE) based Potentiometric Biosensor
A potentiometric biosensor was successfully developed for lead detection in milk
samples. The study was first conducted with free enzyme in synthetic system and a
RESULTS AND DISCUSSION
95
detection limit of 1.93 nanomoles (1.93 μM) was obtained as shown in figure 4.7. A
linear range of lead inhibition was obtained from 1.93 to 4.83 nanomoles (1.93 – 4.83
μM) deduced by change in potential caused by different lead concentrations.
Figure 4.7: Lead inhibition studied on ISE in synthetic system with free enzyme
The method was then applied to nylon immobilized enzyme membrane which was
tied close to the ISE and same method of detection as free system was used.
Hydrosol - gel method of immobilization as mentioned before was carried out and a
detection limit of 9.66 nanomoles (9.66 μM) was obtained in this case. The
percentage inhibition at 9.66 nanomoles (9.66 μM) was 8.27 % and a linear range of
inhibition from 9.66 to 966 nanomoles (9.66 – 966 μM) was observed (figure 4.8).
The relation of percentage inhibition with the lead concentration was found
logarithmic in this case also.
RESULTS AND DISCUSSION
96
Figure 4.8: Lead inhibition studied on ISE in synthetic system with Immobilized
enzyme
The developed potentiometric biosensor was then applied to detect lead in milk
samples without any treatment. The detection limit obtained in this case was also
9.66 nanomoles (9.66 μM) with an inhibition of 9.04 % and linear range of inhibition
from 9.66 to 966 nanomoles (9.66 – 966 μM) as shown in figure 4.9. No data
supporting or contradictory to the present findings is available to validate the results,
still the developed potentiometric biosensor is competent to detect lead equivalent
contamination upto 9.66 nanomoles (9.66 μM). The specificity study was also carried
out in the presence of some masking agents to achieve lead specificity.
RESULTS AND DISCUSSION
97
Figure 4.9: Lead inhibition studied on ISE in milk with Immobilized enzyme
4.5.1 Masking Agent study
The application of masking agents to invalidate the effect of other heavy metals in
the milk was carried out and it was found that EDTA was able to bring the potential
of the milk sample back to the original potential (approximately) in the presence of
all heavy metals by masking as shown by potential shift in table 4.6.
Table 4.6: Urease inhibition study on ISE and EDTA as masking agent for all
heavy metals
Sample Final Potential (mV) Change in potential
(ΔmV)
Control (Milk + 1mM EDTA) -113.6 ± 0.6
Milk + Lead (97 nanomoles) +
All other metals (10
nanomoles ) + EDTA
-114.7 ± 0.3 1.1 ± 0.2
Milk + Lead (97 nanomoles) -116.1 ± 0.4 2.5 ± 0.2
RESULTS AND DISCUSSION
98
Milk + Lead (97 nanomoles) +
All other metals (10
nanomoles)
-118.5 ± 0.2 4.9 ± 0.4
It is evident from the table that 96.6 nanomoles (96.6 μM) of lead decreases the
potential from -113 .6 ± 0.6 mV to -116.1 ± 0.4 mV which further got decreased to -
118.5 ± 0.2 mV in the presence of other heavy metal ions, depicting the increased
inhibition of urease in their presence. It also shows that the inhibition effect of all
heavy metals is more than that of lead alone. The addition of EDTA in the system
lead to masking of all heavy metals as evident from the potential change (-114.7 ±
0.3 mV) which is close to control reaction. Now to mask the affect of all other metals
on urease except lead 1, 10 – Phenanthroline and Sodium potassium tartarate were
used and found to be successful in masking the effect of all other metals present in
the system (table 4.7).
Table 4.7: Masking Agent affect on urease in acid extracted milk
Sample Final Potential
(mV)
Change in
potential (ΔmV)
Control (Milk + 1, 10 – Phenanthroline
and Sodium potassium tartarate)
-98.4 ± 0.4
Milk + Lead (97 nanomoles) + All
other metals (10 nanomoles)
-116.1 ± 0.5 17.7 ± 0.1
Milk + Lead (97 nanomoles) + All
other metals (10 nanomoles) + 1, 10 –
Phenanthroline and Sodium potassium
tartarate
-111.5 ± 0.7 13.1 ± 0.3
Milk + Lead (97 nanomoles) -111.0 ± 0.3 11.6 ± 0.1
RESULTS AND DISCUSSION
99
The table shows that the lead and other heavy metal ions caused the inhibition of
urease and decreased the potential from -98.4 ± 0.4 mV to -116.1 ± 0.5 mV creating
the potential shift of 17.7 ± 0.1 mV which got decreased to 13.1 ± 0.3 mV in the
presence of 1, 10 – Phenanthroline and Sodium potassium tartarate. The potential
drop in the presence of these masking agents was -111.5 ± 0.7 mV which was close
to that caused by lead alone (-111.0 ± 0.3 mV). This similar potential suggest that 1,
10 – Phenanthroline and Sodium potassium tartarate were able to mask the effect of
all other heavy metals except lead and showed the same potential drop as caused by
lead alone.
4.6 Electrochemical Studies
The electrochemical studies were based on direct oxidation or reduction of Pb or Pb2+
respectively on electrode surface.
At low pH Pb 2+
Pb Reduction
pH > 3.5 Pb Pb 2+
Oxidation
4.6.1 Lead detection on platinum working electrode
The lead detection on platinum electrode was done at low pH (0.9 – 1.2) as well as
on neutral pH. The low pH studies were done to configure whether the developed
method could be applied to acid extracted milk samples directly (without adjusting
the pH to neutral) or not. It was first optimized on synthetic system and then applied
to acid extracted milk samples. The parameters of the study are mentioned in the
figure 4.11. In case of synthetic system at low pH the linear range of lead detection
was found to be from 4.8 X 10-2
nanomoles to 4.8 micromoles (4.83 nM to 483 μM)
as shown in figure 4.10 with lower limit detection of 4.8 X 10-2
nanomoles (4.83
RESULTS AND DISCUSSION
100
nM). Tsai et al. (2001) has achieved a detection limit of 0.1 µM using boron doped
diamond electrode and a detection limit of 48.3 nM has been achieved by Babyak
and Smart (2004) using the same electrode. Goubert – Renaudin et al. (2009) used
amide – cyclan –functionalized silica modified CPE to attain 2.7 nM detection limit.
Figure 4.10: Lead detection in synthetic system ( pH 0.9-1.2). b1.bin - 0.1 M KCl,
p1.bin - 4.8 X 10-2
nanomoles Pb2+
, p2.bin – 4.8 X 10-1
nanomoles Pb2+
, p3.bin -
4.8 nanomoles Pb2+
, p4.bin - 48 nanomoles Pb2+
, p5.bin – 480 nanomoles Pb2+
,
p6.bin - 4.8 micromoles Pb2+
Locally available verka milk samples were used for lead spiking studies. In case of
acid extracted milk the linear range was observed from 4.8 X 10-2
nanomoles to 4.8
micromoles (4.83 nM – 483 μM) Pb2+
as shown in figure 4.11 with detection limit of
4.8 X 10-2
nanomoles (4.83 nM). The method was successfully applied to the milk
samples but the linear range was narrowed down to 48 nanomoles (4.83 μM).
RESULTS AND DISCUSSION
101
Figure 4.11: Lead detection in acid extracted milk (pH 0.9-1.2). b2.bin – 0.1 M
KCl, p1.bin - 4.8 X 10-2
nanomoles Pb2+
, p2.bin – 4.8 X 10-1
nanomoles Pb2+
,
p3.bin - 4.8 nanomoles Pb2+
, p4.bin - 48 nanomoles Pb2+
From the current values of different concentration of lead in synthetic and acid
extracted milk sample (table 4.8), a comparative relation between the two systems
was developed which could be applied to detect lead concentration in unknown milk
samples.
Table 4.8: Comparison of results of synthetic system and milk samples at low pH
Lead Conc
(nanomoles)
Current in
Synthetic
system (
10-1
mA)
Change in
current in
synthetic
system (∆I)
Current in
milk
samples (
10-1
mA)
Change in
current in
milk
samples
(∆I)
Change
difference
in synthetic
and milk
samples
Blank (0.1 M
KCl)
1.079 - 1.925 (milk
w/o spiking)
- 0.846
RESULTS AND DISCUSSION
102
4.8 X 10-2
1.134 0.055 2.109 0.184 0.975
4.8 X 10-1
1.195 0.116 2.258 0.333 1.063
4.8 1.297 0.218 2.350 0.425 1.053
48 1.386 0.307 2.442 0.517 1.056
From the table it could be furnished that in case of milk an addition of 4.8 X 10-2
nanomoles
(4.83 nM) Pb2+
leads to a difference of 0.129 units in current (0.975 –
0.846 = 0.129) with respect to control which corresponds to a contamination of
approximately 4 X 10-1
nanomoles (48.3 μM ).
Lead detection at neutral pH
Lead detection on neutral pH was done on synthetic samples first and it was observed
that, only the detection of 0.48 micromoles (48.3 μM) lead and above was possible as
shown in figure 4.12. The detection limit (0.48 micromoles or 4.83 μM) and linear
range (0.48 – 4.8 micromoles Pb2+
or 48.3 – 483 μM) was found very much higher
than the permissible limit of lead in milk (96.6 nM) so the study was further not
applied to milk samples. The neutral pH also leads to fast fouling of the platinum
electrode surface and hence loss of sensitivity. So the electrode surface needs to be
regularly polished after few reactions. This polishing step adds time to the whole
process and still was not able to decrease down the detection limit below 0.48
micromoles Pb2+
(48.3 μM).
RESULTS AND DISCUSSION
103
Figure 4.12: Lead detection in synthetic system at neutral pH. b2.bin- 0.1 M
KCL, Pb10.bin- 0.48 micromoles Pb2+
, p100.bin- 4.8 micromoles Pb2+
4.6.1.1 Preconcentration study on platinum working electrode
For preconcentration of lead on platinum electrode an alginate- sol gel hybrid
membrane was developed. The alginate-TEOS hybrid was able to form a thin
membrane capable to be used as adsorption matrix on electrode tip. The SEM studies
of the membrane surface (figure 4.13) reveal microstructures showing alginate-
silicate interactions that results in small vesicles with increased surface area which
forms the platform for metal adsorption.
Figure 4.13: SEM image of alginate- TEOS hybrid membrane
RESULTS AND DISCUSSION
104
Sodium alginate – sol gel membrane was used for the preconcentration of lead on the
platinum electrode. It was observed that the membrane was able to preconcentrate the
lead as the current was found to be decreased (suppression of current) after applying
the membrane (figure 4.14) but the release of lead complex from the membrane was
not achieved.
a
b
Figure 4.14: Suppression of Lead peak after preconcentration. A- 0.48
micromoles lead without membrane, b- 0.48 micromoles lead with membrane
Promising results were obtained in heavy metal preconcentration studies of the
membrane. As per the electrochemical studies, the membrane applied at the tip of the
electrode was able to preconcentrate almost 99% lead from the solution. The
oxidation peak of 0.48 micromoles (48.3 μM) lead got completely suppressed after
the application of the membrane as shown in the figure 4.15 and as visualized after
reaction the membrane’s morphology changed after complexation with the metal.
Affinity of alginic acid towards metal cations like lead has already been postulated by
Jeon et al. (2002a) and Jeon et al. (2002b). The lead adsorption efficiency has been
improved using magnetically modified alginic acid by Jeon et al. (2007). The
observed interaction of the membrane with lead is also advantageous in developing
RESULTS AND DISCUSSION
105
urease immobilized membrane for heavy metal detection as the membrane will be
able to concentrate cations close to the enzyme and hence enhanced inhibition could
be observed towards heavy metal cations. Although this postulated theory was not
further extended in practical work as the release of lead from the membrane was not
achieved successfully. Polarization of Lead- EDTA complex studied on all kind of
electrodes – Platinum, Glassy carbon, screen printed was studied but no distinction in
different concentration of lead was observed. As per the studies conducted, it is
suggested that the prepared alginate-TEOS sol gel membrane is a very promising
matrix for heavy metal preconcentration and its removal from the wastewaters.
4.6.2 Lead detection on Glassy Carbon Electrode (GCE)
Reduction studies of lead were also carried out on Glassy carbon electrode as the
working electrode and a similar pattern of reduction as platinum electrode was
observed as shown in figure 4.15.
Figure 4.15: lead detection on GCE. KCL5- 0.1 M KCl, g3- 4.8 X 10-2
nanomoles
Pb2+
, g6- 4.8 X 10-1
nanomoles Pb2+
, g9- 4.8 nanomoles Pb2+
, g12- 48 nanomoles
Pb2+
, g14- 480 nanomoles Pb2+
, g16- 4.8 micromoles Pb2+
RESULTS AND DISCUSSION
106
The linear range for lead was found to be from 4.8 X 10-2
nanomoles to 4.8
micromoles (4.83 nM – 483 μM) Pb2+
, but the curves were not sharp and as smooth as
on platinum electrode, so platinum electrode was assumed to be better for direct
estimation of lead by reduction.
4.6.3 Lead detection on mercury coated Glassy Carbon Electrode (GCE)
After getting a very low detection limit on platinum electrode and unmodified glassy
carbon electrode, the study was extended to mercury coated GCE to make the study
possible at neutral pH with low detection limits. For this the bare GCE was coated
with mercury and then the polarization of lead (483 μM) was done. It was observed
that the lead polarization took place at specific potential with a very distinct lead peak
at -0.6 V. To sharpen the peak and to make it more stable different parameters were
optimized.
4.6.3.1 Effect of pH on lead polarization
Lead polarization studies from pH 1 to 7 were done and it was observed that pH 1 to 3
does not result in any lead peak at - 0.6V (figure 4.16). There was a very prominent
peak at pH 4 which got reduced at higher pH values. So pH 4 was taken as to be the
optimum pH for lead reduction on mercury coated GCE. A pH range of 4 -5 was also
optimized by Kadara and Tothill, 2004 for lead detection on bismuth film screen
printed electrode.
RESULTS AND DISCUSSION
107
Figure 4.16: Effect of different pH on lead reduction
An electrochemical sensor was developed by Bouwe et al. (2011), for sub- nanomolar
detection of lead (0.4 nM). A 1,10- phenanthroline – montmorillonite intercalate was
formed at acidic and neutral pH and was used as Pb2+
sensor via CPE by adsorptive
stripping voltametry. The optimal pH found in this case was 6.
4.6.3.2 Effect of different buffers on lead polarization
0.1 M KCL, 0.1 M Acetate buffer and 0.05 M Tartarate buffer were used for lead
reduction. No peak corresponding to 483 µM lead was observed in acetate and
tartarate buffer and only KCl buffer was able to deliver a sharp peak at -0.6 V
potential (figure 4.17). So for further study KCl buffer was used for carrying out lead
reduction.
RESULTS AND DISCUSSION
108
Figure 4.17: Effect of different buffers on lead reduction. p15.bin – 0.1 M KCl,
p6.bin – Acetate buffer, p8.bin – Tartarate buffer
4.6.3.3 Effect of KCl concentration on lead polalization
From the different buffers used for lead reduction it was very clear that KCl was the
most excellent buffer for carrying out lead reduction. So the effect of KCL
concentration was studied on current enhancement of lead peak as shown in figure
4.18. It was observed that there was an increase in current value of lead from 0.1 to 1
M KCl. At 2 M KCL the current got drastically dropped as compared to the 1M KCL.
So as an optimum value 1 M KCL was used in further study.
Babyak and Smart 2003, had used 0.1 M KCl as supporting electrolyte for lead
detection on boron doped diamond electrode using square wave anodic stripping
voltametry. Hassan et al. (2008) had optimized 0.1 M KCl as supporting electrolyte
for anodic stripping voltametric detection of Pb2+
using carbon paste electrode
modified with chitosan.
RESULTS AND DISCUSSION
109
Figure 4.18: Effect of different concentrations of KCl buffer on lead reduction
After the optimization of all the parameters lead detection was done on mercury
coated GCE and a linear range from 0.96 to 3.8 micromoles (96.6 – 386 μM) lead was
obtained (figure 4.19). A drastic increase in current at 4.8 micromoles (483 μM) lead
was observed and the lowest limit of detection obtained in this case was 0.48
micromoles (48.3 μM).
Figure 4.19: Lead detection on mercury coated GCE. p2.bin – 0.97 micromoles
Pb2+
, p5.bin – 1.93 micromoles Pb2+
, p7.bin – 2.9 micromoles Pb2+
, p9.bin – 3.8
micromoles Pb2+
, p11.bin – 4.8 micromoles Pb2+
A detection limit of 1.93 X 10-1
nM was achieved by Liu et al. (2005), using carbon
nantotubes- nanoelectrodearrays (CNTs-NEAs) coated with bismuth film. In a similar
RESULTS AND DISCUSSION
110
study as ours, Loh et al. (2006), achieved a detection limit of 4.83 µM using screen
printed electrode (SPE) with poly(vinyl) chloride (PVC) as printed matrix. Lead was
deposited with mercury on working electrode at a pH of 2. The various parameters
optimized were, deposited time = 15sec, deposited potential = 1000 mV, scanning
range = 20mV sec-1
and amplitude = 25. Linearity obtained in this case was from 4.83
to 242 µM of Pb2+
. Hassan et al. (2008), has achieved a detection limit of 9.66 nM
using carbon paste electrode modified with chitosan. Later Senthilkumar and
Saraswathi (2009) developed an electrochemical sensor for lead detection using a
zeolite NH4- Y modified carbon paste electrode (ZYMCPE) with a detection limit of
17.3 µM. Recently enhanced anodic stripping Voltammetry method has been used for
blood lead determination (Zyoud and A-Subu, 2011)
4.6.4 Development of Electrochemical Biosensor using Carbon Paste Electrode
(CPE)
The electrochemical study was conducted on CH Instruments Electrochemical
workstation, Model 660. The working electrode used was the self fabricated carbon
paste electrode with Ag/AgCl2 as reference electrode and platinum wire as the counter
electrode. The oxidation of NADPH was studied in the reaction catalyzed by urease
and glutamate dehydrogenase GLDH present in the whole cell of B. sphaericus. The
study was based on the detection of unspent NADPH after the reaction of urease on
urea and GLDH on α- Ketoglutarate as shown in the Eq 1 & 2.
Urea + H2O HCO3- + 2 NH4
+ (1)
2 NH4+ + α- Ketoglutarate + NADPH + H
+ L-Glutamate + NADP
+
(2)
GLDH
Urease
RESULTS AND DISCUSSION
111
The ammonia produced by action of urease on urea was taken up by α- Ketoglutarate
added exogenously in the system in the presence of NADPH and resulted in the
production of NADP+. As the reaction would not be able to oxidize whole of the
NADPH present in the system, the oxidation of unoxidised NADPH left after the
reaction was studied electrochemically and this was correlated with the lead
concentration. The increase in lead concentration caused the inhibition of urease and
resulted in more amount of unspent NADPH. This caused increase in the current of
the NADPH oxidation as compared to the lead deficient system and formed the basis
of the study. To achieve proper NADPH oxidation various parameters were optimized
and then the lead inhibition studies were conducted on milk as such and after
preconcentration through lead specific column.
4.6.4.1 Optimization of NADPH concentration
The first step in the study was to optimize the minimum amount of NADPH required
to obtain the detectable NADPH oxidation peak. Oxidation of NADPH was studied
amperometrically in a range of - 0.2 to 0.8 V and an oxidation peak was obtained at
0.1V- 0.15V. The current value of different CPE carrying same enzyme units and
prepared at the same time was - 4.4662 mA, with standard deviation of ± 0.784 A. A
significant shift in potential towards higher side was observed with increase in
NADPH concentration. There was a linear increase in current with increasing
NADPH concentration from 4.3 x 10-2
mM to 1.3 x 10-1
mM (figure 4.20), beyond
which fouling of the electrode was observed. The optimized NADPH which has been
added exogenously is 4.3 x 10-2
mM. In a similar study by Rodriguez et al. (2004a), a
fixed concentration of 0.283 mM NADH has been used, but the required amount has
been reduced to very low levels in the present study, owing to the use of whole cell
RESULTS AND DISCUSSION
112
instead of enzyme. Azmi et al. (2009) has also showed that a level of 5.0 x 10-2
mM
NADH is detectable using simple UV-Vis spectrophotometer, which could be further
lowered down by sensitive detecting system as obtained in the present study using
electrochemical workstation.
Figure 4.20: Cyclic voltametric studies on electrochemical oxidation of NADPH.
c7.bin: 4.3 x 10-2
mM NADPH, c8.bin: 8.6 x 10-2
mM NADPH, c9.bin: 1.3 x 10-1
mM NADPH
4.6.4.2 Reaction time optimization
In case of reaction time, an oxidation peak at 0.15 V was observed at 3 and 5 minutes
incubation with enhanced current at 5minutes incubation (figure 4.21). The 10 & 20
minutes incubation gave less and almost same current along with potential shift
towards higher side as compared to the 5 minutes incubation. So the response time
optimized for the reaction is 5 minutes, preceded by an incubation of electrode with
lead for 10 minutes as done by other workers (Ilangovan et al., 2006) to increase the
inhibition efficiency. The same microbe has been shown to give response against
nickel ions in 1.5 minutes (Verma and Singh, 2006).
RESULTS AND DISCUSSION
113
Figure 4.21: NADPH oxidation after different reaction times
4.6.4.3 Temperature effect on NADPH oxidation
The temperature range from 20˚C to 45 ˚C was studied. It was observed that, almost
same amount of current was generated at 20 & 25 ˚C, showing same amount of
oxidizable NADPH availability (figure 4.22). At higher temperatures (30˚C, 37 ˚C &
45 ˚C) there was an increase in current and a potential shift towards lower side was
observed, depicting the inability of the system to oxidize the excess NADPH
completely. At 37 ˚C & 45 ˚C, the oxidation peak also got broadened, depicting the
non- suitability of this temperature for the reaction. So a temperature range of 20 – 25
˚C was optimized for the present study.
Figure 4.22: Temperature effect on NADPH oxidation
RESULTS AND DISCUSSION
114
4.6.4.4 Biomass Optimization
The optimum biomass loading was obtained by immobilizing different biomass
concentrations, with different enzyme units at the tip of the electrode. A study was
conducted in which biomass containing 0.975, 1.95 & 2.915 enzyme units of urease
was immobilized in the CPE and NADPH oxidation was recorded. The increase in
biomass increases the amount of intracellular NADPH available, giving the increase
in current. A sequential higher side potential shift with increasing enzyme units was
observed (figure 4.23). The oxidation peak got broadened at 2.915 U. Inhibition
studies of lead was conducted on all the biomass concentrations and it was observed
that the inhibition affect of lead was less pronounced at higher biomass loading,
which could be because of overloading of biomass. The biomass containing 1.95 U of
urease having a current value of - 4.800 ± 0.313 A, was optimized for oxidation
studies, as inhibition effect of lead was less pronounced at 0.975 and 2.915 U. These
enzyme units are much less as compared to 5 U used by Rodriguez et al. (2004a), in
different immobilization methods for urease. Lesser enzyme units used in the study
indicate better kinetic properties of the immobilized biomass within the carbon paste
without much loss in enzyme activity. The whole cell immobilized CPE was found to
be disposable in nature.
RESULTS AND DISCUSSION
115
Figure 4.23: Biomass optimization for NADPH oxidation. c2.bin: 0.975 U, c3.bin:
1.95 U, c4.bin: 2.915 U
4.6.4.5 Lead Inhibition and monitoring studies
The inhibitory studies for lead were conducted in a concentration range of 4.8 X 10-4
to 0.48 nanomoles (4.83 nM – 4.83 μM) Pb2+
. The linear range of inhibition was
observed from 4.8 X 10-4
to 4.8 X 10-2
nanomoles (4.83 – 483 nM) lead in synthetic
samples (figure 4.24), with shift in potential with increasing lead concentration. The
current at 0.48 nanomoles (4.83 μM) Pb2+
got drastically reduced owing to the
electrode fouling due to excessive NADP+ production. A logarithmic relationship was
observed between current value and lead concentration (figure 4.25). Lower limit of
detection achieved was 4.8 X 10-4
nanomoles (4.83 nM) Pb2+
.
RESULTS AND DISCUSSION
116
Figure 4.24: Inhibition study of lead in synthetic system. c1.bin: 4.8 X 10-4
nanomoles Pb2+
, c2: 4.8 X 10-3
nanomoles Pb2+
, c3: 4.8 X 10-2
nanomoles Pb2+
, c4:
4.8 X 10-1
nanomoles Pb2+
Figure 4.25: Calibration curve of Pb2+
ions
Milk as such and spiked milk samples were analyzed for lead contamination by two
different ways. One is milk samples without any pretreatment and the other, after acid
extraction pretreatment. In direct milk samples there was a large shift in potential at
all lead concentrations in comparison to synthetic samples that could be because of
delayed NADPH oxidation as a result of interference by milk proteins and other milk
constituents. Linear range of detection of 4.8 X 10-3
nanomoles to 4.8 X 10-1
nanomoles (4.83 nM – 4.83 μM) Pb2+
was obtained in this case (figure 4.26) and
lower limit of detection was 4.8 X 10-4
nanomoles (4.83 nM) Pb2+
.
RESULTS AND DISCUSSION
117
Figure 4.26: Linear range detection of Pb2+
ions in Milk. c2.bin: 4.8 X 10-3
nanomoles Pb2+
, c3.bin: 4.8 X 10-2
nanomoles Pb2+
, c4.bin: 4.8 X 10-1
nanomoles
Pb2+
In case of acid extracted milk, lead specificity was achieved by passing the milk
sample through lead selective column (Pb. SpecTM
), which showed a percentage
loading of 79.26 ± 0.8 and a recovery of 85.44% ± 0.4 Pb2+
as detected through
Sodium dithiocarbamate method using the following calibration curve (figure 4.27).
Figure 4.27: Calibration curve of lead using Sodium Dithiocarbamate
In case of acid extracted milk, the linear detection range was from of 4.8 X 10-4
nanomoles to 4.8 X 10-1
nanomoles (4.83 nM – 4.83 μM) Pb2+
(figure 4.28) and a
RESULTS AND DISCUSSION
118
lower limit detection of 2.4 X 10-4
nanomoles (2.4 nM) was achieved. The percentage
inhibition in milk was less as compared to the synthetic samples. It could be defended
by two reasons, firstly the intricate matrix of the milk complexes with the lead and
reduces it inhibition efficiency, secondly the passage of the lead containing milk
sample through the lead column leads to some loss of the lead as the percentage
loading and the recovery from the column is not 100 %. The reliability of the
developed biosensor was checked by studying the change in current (ΔI) of lead
spiked synthetic sample and acid extracted milk samples. The ΔI values obtained for
synthetic and milk sample was 0.949 ± 0.109 and 1.011 ± 0.021 A respectively, which
are comparable. Lead analysis was done in spiked milk samples with the developed
biosensor and the results obtained were as given in table 4.9.
Figure 4.28: Linear range detection of Pb2+
ions in Acid extracted Milk. c1.bin:
4.83 X 10-4
nanomoles Pb2+
, c3.bin: 4.83 X 10-2
nanomoles Pb2+
, c4.bin: 4.83 X 10-
1 nanomoles Pb
2+
Table 4.9: Reliability studies of the electrochemical biosensor
Lead added (nanomoles) Lead found (nanomoles)
4.8 X 10-4
4.7 ± 0.113 X 10-4
48 X 10-3
4.76 ± 0.483 X 10-3
RESULTS AND DISCUSSION
119
The developed biosensor was applied for lead analysis in various milk samples from
different places and a maximum of 48 X 10-3
nanomoles (48.3 nM) lead
contamination was found, most of the samples were found to be safe for human
consumption.
Finally, lower limit of detection (LOD) for lead ions has been compared with
biosensors developed by other workers in table 4.10. It appears that the present
biosensor has far more better detection limit for lead than developed so far. Fungal
and Plant ureases are hexameric (α6) whereas bacterial ureases are heterotrimeric
(αβγ)3. The Bacillus sphaericus is a novel isolate with better kinetic properties than
jack bean and Bacillus pasteurii (Verma and Singh, 2003), hence gives lower limit of
detection for heavy metals (Verma and Singh, 2006). The uncompetitive inhibition of
urease by lead suggests that unlike Cu2+
and Ag2+
which bind to nitrogen of histidine
and possibly oxygen of aspartic and glutamic acid residues in the active site of urease
(Saboury et al., 2010), lead binds to thiol group of cysteine residues present in the flap
which causes modification in cysteine residues due to which the flap loses its mobility
and activity of urease is inhibited (Krajewska and Zaborska, 2007). Biosensors
developed so far had an application of monitoring lead ions in water samples only, but
the present biosensor developed has an application in drinking water as well as in
milk samples, which is a first endeavor of its kind for milk lead analysis at such a low
levels. Babkina and Ulakhovich (2004) had used ssNDA to develop an amperometric
biosensor for heavy metal analysis in food samples and attained a detection limit of in
milk, the authors work was able to achieve the detection limit of 2.4 X 10-4
nanomoles
in milk which is under the permissible limit of lead in milk.. Moreover, the present
biosensor is also very simple, economical and more sensitive than the DNAzyme
based biosensors (Li and Lu, 2000; Xiao et al., 2007; Shen et al., 2008) and the lead
RESULTS AND DISCUSSION
120
specificity has been achieved using the lead specific column, making it a complete
lead specific biosensor.
Table 4.10: Comparison between limits of detection (LOD) of Pb2+
ions
with different biosensors
Biocomponent used Transducer Pb (II) LOD Reference
L-Lactate
dehydrogenase
Amperometrer 0.2 µM Fennouh et al. (1998)
Cysteine
desulfhydrolase
Amperometer 20 nM Kremleva et al. (1999)
Alkaline phosphatase Optical 0.1 nM Veselova and
Shekhovtsova (2000)
DNAzyme Optical 10 nM Liu and Lu (2000)
Antibody Optical 6 nM Blake et al. (2001)
Urease Conductometer 0.9 mM Lee and Lee, (2002)
Cyanobacteria Amperometer IC50 : 1.6 µM Tay et al. (2003)
Urease Optical 1 X 10-2
mM Kuswandi (2003)
Urease Optical 0.1 mM Tsai et al. (2003)
ssNDA Amperometric 0.1 nM Babkina and Ulakhovich
(2004)
Urease Electrochemical 1 mM Ogonczyk et al. (2005)
Human Angiotensin I Electrochemical 1.9 nM Chow et al. (2005)
Urease and
Acetylcholineesterase
Optical 4.83 nM Haron and Ray (2006)
Urease Conductometer 1 mM Ilangovan et al. (2006)
RESULTS AND DISCUSSION
121
DNAzyme Electrochemical 0.3 µM Xiao et al. (2007)
Invertase Electrochemical 30 nM Bagal-Kestwal et al.
(2008)
Alkaline phoaphatase Conductometer
1.93 x 10-2
µM
Berezhetskyy et al.
(2008)
dsNDA Electrochemical 2 mM Oliveira et al. (2008)
DNAzyme Electrochemical 1 nM Shen et al. (2008)
Urease (Bacillus
sphaericus)
Electrochemical 2.4 nM Present study
4.7 Development of DNAzyme based Optical Biosensor
The DNAzyme interaction with lead which results in breakdown of substrate
strand has been taken as the basis of this study. To carry out the fluorescence read
microarray plate reader was first calibrated using fluorescein as the fluorescent
indicator. Different protocols were tried to construct a standard curve for known
concentrations of fluorescein. The final protocol developed was as follows:
Protocol for fluorescence read
Incubate at 30 ˚C for 30 sec
Plate read
Go to line 1 for 2 more times
END
Lid Temperature- Off
Lid off temperature – 30 ˚C
Volume set as used
RESULTS AND DISCUSSION
122
The standard curve for fluorescein obtained shown in figure 4.29
Figure 4.29: Standard curve of Fluorescein obtained at plate reader
4.7.1 DNAzyme study in free system
In initial study 17 E (donor) labeled with fluorescein (Ex = 494 nm, Em = 521
nm) and 17DS labeled with TAMRA (Ex = 542 nm, Em = 568 nm ) were first
hybridized according to Swearingen et al. (2005). The donor (17 E) was excited at
494 nm (λ1) and read against sybergreen (Ex = 497 nm, Em = 520 nm) in the plate
read. The acceptor (17 DS) was excited at 542 nm (λ2) and its emission was read at
568 nm (λ3) against HEX (Ex = 535 nm, Em = 556 nm) in the plate reader as shown
in the table 4.11. The hybridization was confirmed by taking the fluorescence read of
both the strands separately and after the hybridization, shown in table 4.12.
RESULTS AND DISCUSSION
123
Table 4.11: Donor – Acceptor Couple
Designation Wavelength Relation
λ1 494 nm Excitation of donor (fluorescein)
λ2 542 nm Emission of donor/ Excitation of
acceptor (TAMARA)
λ3 568 nm Emission of acceptor (TAMARA)
Table 4.12: Fluorescence read for hybridization
Reagents Fluorescence Wavelengths
Donor 17 E -
Fluorescein
0.812 ± 0.107 Ex at λ1, Em at λ2
Acceptor 17 S - TAMRA 0.134 ± 0.035 Ex at λ2, Em at λ3
Hybrid 0.468 ± 0.037 Ex at λ1, Em at λ3
As the 17 E was acting as donor and 17DS as acceptor, it was observed that
after hybridization the fluorescence of acceptor got enhanced from 0.134 ± 0.035 to
0.468 ± 0.037 supporting the phenomenon of FRET as occurrence of FRET causes
transfer of energy from donor to acceptor thus increasing its fluorescence. After the
successful attainment of hybridization DNAzyme interactions with Pb2+
was studied.
It was observed that the addition of Pb2+
leads to decrease in fluorescence intensity of
acceptor (17 DS) suggesting that it is catalyzing the cleavage of substrate strand by
the enzyme strand and hence the break of FRET also (table 4.13).
RESULTS AND DISCUSSION
124
Table 4.13: DNAzyme - Pb2+
interaction
Reagents Fluorescence
(λ3)
Hybrid + water 0.490 ± 0.011
Hybrid + Lead (0.48 nanomoles) 0.417 ± 0.022
Hybrid + Lead (0.72 nanomoles) 0.402 ± 0.012
Till 2005 the DNAzyme based lead biosensors were constructed using a
fluorophore and a quencher combination (mainly TAMRA as fluorophore and Dabcyl
as quencher) in which the hybridization leads to quenching of the fluorescence and
cleavage of substrate strand leads to increase in fluorescence (Liu, 2002; Chang et al.,
2005; Swearingen et al., 2005; Wernette et al., 2006; Shen et al., 2007). The
efficiency has been increased by using two quenchers, one at the 3’substrate strand
and other at 3’enzyme strand. A fluorophore at 5’ substrate strand gets quenched by
both inter and intra- molecular quenchers decreasing the backgroung fluorescence
(Liu and Lu, 2003). Zhang et al. (2010a) has utilized the potential of DNAzyme as
catalytic and molecular beacon to achieve a detection limit of 0.6 nM. The substrate
strand has been modified to form an intermolecular beacon which has a fluorophore at
one end and a quencher at the other. The cleavage of the substrate strand leads to the
separation of the quencher and the fluorophore and hence increase in the fluorescence.
Now a new era of DNAzyme based sensors has arisen in which new molecular probes
(G rich oligonucleotides and G quadruplex) has been optimized with some
modifications to give better signals with Pb2+
ions. Li et al. (2010a) has developed a
Pb2+
ion detection system based on peroxidase like activity of K+ stabilized G
quadruplex DNAzyme named PS2.M with two detection means. The K+ stabilized
RESULTS AND DISCUSSION
125
PS2.M binds to hemin with high efficiency forming a complex that mimics
horseradish peroxidase (HRP) and catalyzes H2O2 mediated oxidation of 2,2’ – azino
– bis (3 – ethylbenzothiazoline-6 – sulfonic acid) diammonium salt (ABTS) and
luminal to generate color change or chemiluminiscence (CL) emission. But the
addition Pb2+
ion leads to some conformational change in the PS2.M quadruplex
which then does not bind to hemin. This conformational transition is accompanied by
decrease in DNAzyme activity and therefore less change in color and CL emission.
Pb2+
ion could be detected upto 33.8 nM colorimetrically and 9.66 X 10-4
nM using
the CL emission method. Later the same workers, Li et al. (2010b) constructed a
novel turn – on fluorescent sensor based on cleavage of T30695 and X duplex in the
presence of Pb2+
ions and formation of a G4 structure stabilized by Pb2+
ions. The
Pb2+
ions stabilized quadruplex could be detected using Zn protoporphyrin IX
fluorescent probe. This duplex – quadruplex exchange system has a limit of detection
of 20 nM. In an ease to develop a more sensitive sensor Li et al. (2011) has now
developed a sensor in which Pb2+
ions induces formation of G4 quadruplex of
AGRO100 molecule that can bind to hemin to form complexes. This G4 AGRO100 –
hemin complex can effectively catalyze the H2O2 mediated oxidation of amplex
UltraRed (AUR), resulting in increase in fluorescence. The detection limit achieved
by this device was 0.4 nM. The present study is first one which has used a donor
acceptor based FRET method for constructing the lead specific biosensor. Recently,
Ma et al. (2011) developed an electrogenerated chemiluminiscence (ECL) biosensor
based on the use of ruthenium complex tagged 5’- amino -17 E’ as an ECL probe. The
modified Ru 1-17 E’ and substrate strand were covalently coupled on a graphite
electrode modified with 4 – aminobenzoic acid. Pb2+
ions induced cleavage of
RESULTS AND DISCUSSION
126
substrate strand leads to dissociation of double stranded DNA complex and hence
high ECL signals. The detection limit achieved by the workers is 1.4 pM.
4.7.2 DNAzyme study in hydrosol- gel immobilization matrix
The immobilization of hybridized DNAzyme in hydrosol- gel matrix was achieved
using TMOS and TEOS solvents. It was observed that immobilization of DNAzyme
in TMOS leads to better gelatin of the mixture in the well and more fluorescence
intensity compared to control (system without DNAzyme) and TEOS as shown in
table 4.14.
Table 4.14: Fluorescence with different immobilization matrices
Solvent Fluorescence Intensity
(λ3)
TMOS + Water 0.406 ± 0.024
TEOS + Water 0.331 ± 0.050
But the fluorescence change of hybridized DNAzyme in immobilized state was low
compared to the free system as compared to table 4.13, similar to the observations of
Shen et al. (2007). The signal development efficiency of the DNAzyme couple is also
dependent on the sequence of both the strands and the various modifications done to
them (Shen et al., 2007). Also the production of methanol as a byproduct of silane
hydrolysis do not lead to any loss of DNAzyme function (Shen et al., 2007), making it
a desirable method of immobilization for DNAzyme based systems. The detection
limit obtained in immobilized system was 0.12 nanomoles (24 μM) and an appropriate
decrease in fluorescence was observed with higher amount of Pb2+
ions due to break
RESULTS AND DISCUSSION
127
in FRET phenomenon (table 4.15). As the detection limit achieved was higher further
application of the developed method for milk sample was not studied.
Table 4.15: Lead interaction with Hydrosol- gel immobilized DNAzyme
Reagents Fluorescence
(λ3)
Hybrid + TMOS + Water 0.406 ± 0.024
Hybrid + TMOS + 0.12
nanomoles (24 μM) lead
0.294 ± 0.004
Hybrid + TMOS + 0.24
nanomoles (483 μM) lead
0.265 ± 0.025
4.7.3 DNAzyme study on gold chip
DNAzyme was successfully immobilized on gold chip according to Ananthanawant et
al. (2000). The usual protocol for florescence read was modified to carry out the
initial 70 ˚ C incubation of the 17 E strand on the gold chip as mentioned below.
Protocol for high temperature incubation and
fluorescence read
Incubate at 70 ˚C for 20 mins
Plate read
Go to line 1 for 2 more times
END
Lid Temperature- 70 ˚C
Lid off temperature – 30 ˚C
Volume set as used
RESULTS AND DISCUSSION
128
The immobilization of 17 E strand on gold surface was confirmed by first taking the
fluorescence of bare gold chip and then after each step, immobilization of 17 E strand
and during hybridization of 17 E and 17 DS (table 4.16).
Table 4.16: Fluorescence read after each step of immobilization and
hybridization on gold surface
Steps Fluorescence read
(λ3)
Bare gold chip 0.075 ± 0.010
After 17 E - TAMRA immobilization
(acceptor) and fixation with mercaptoethanol
0.146 + 0.015
After hybridization with 17 DS –Fluorescein
(donor)
0.271 ± 0.043
In this case 17 DS was labeled with fluorescein which acted as donor and 17 E strand
with TAMRA which acted as acceptor to create FRET phenomenon. To take the
fluorescence of both the strands separately HEX (Em = 556 nm) was used as the read
out dye for 17 E and Sybergreen (Em = 520 nm) for 17 DS. The increase in
fluorescence of bare gold surface after 17 E immobilization indicates the presence of
fluorescently labeled moiety on gold surface. The treatment with mercaptoethanol
fixes the enzyme strand and then washing lead to the removal of unbound enzyme
strands as evident from the fluorescence read after washing (table 4.16). The
hybridization was achieved successfully as it leads to the increase in fluorescence of
the 17 E (acceptor) strands to almost double.
RESULTS AND DISCUSSION
129
After hybridization DNAzyme was interacted with Pb2+
ions and volume was
miniaturized upto 5µl volume on the gold chip. The change in fluorescence observed
with Pb2+
ions in this system was very low as shown in the table 4.17.
Table 4.17: Lead interaction with DNAzyme immobilized on gold surface
Reagents Fluorescence (F)
At λ3
Change in
Fluorescence (ΔF)
Control 0.166 ± 0.113
0.06 nanomoles Lead (5µl
of 12 μM)
0.124 ± 0.110 0.042 ± 0.003
0.12 nanomoles Lead (5µl
of 24 μM)
0.112 ± 0.109 0.054 ± 0.004
As the change in fluorescence observed was very low enzyme and substrate
strand concentration was optimized to get more signal enhancement and lower limit of
detection. Firstly the 17 DS was optimized for maximum concentration and it was
observed that 1 µM and 2 µM 17 DS had almost same fluorescence (table 4.18), so 1
µM 17 DS was optimized for further experiments.
Table 4.18: Substrate (17 DS) strand optimization
17 DS Concentration
(µM)
Fluorescence at λ3
1 1.338 ± 0.009
2 1.321 ± 0.005
3 1.670 ± 0.177
RESULTS AND DISCUSSION
130
Secondly different concentrations of enzyme strand (17 E) were used for
hybridized DNAzyme and change in fluorescence was observed (table 4.19).
Table 4.19: Enzyme (17 E) strand optimization
17 E Conc.
(µM)
Fluorescence
with water at λ3
Fluorescence
with 0.24
nanomoles Pb2+
ions at λ3
Change in
Fluorescence (ΔF)
1 0.326 ±0.018 0.263 ± 0.026 0.063 ± 0.008
2 0.715 ± 0.023 0.109 ± 0.142 0.606 ± 0.119
3 0.660 ± 0.045 0.315 ± 0.146 0.315 ± 0.101
As maximum change in fluorescence was observed with 2 µM 17 E, it was
optimized for further experiments. Also the effect of Pb2+
ions was also more
pronounced at 2 µM. A decrease in fluorescence at 3 µM 17 E concentrations could
be attributed to higher dye content which failed to follow the Lambert’s Beer Law.
As indicated by the observations 1 µM 17 DS and 2 µM 17 E was finally
optimized based upon their fluorescence signal to carry out lead interaction study. In a
similar study a concentration of 1 µM 17 E and 17 DS was used by Swearingen et al.
(2005) to develop lead biosensor after immobilization on gold surface. The present
work has utilized the advantage of DNAzyme immobilization on gold surface through
thiolation of enzyme strand as done by Swearingen et al. (2005). This allowed better
sensitivity and lead analysis on solid surface which decreases background signals.
Wernette et al. (2006) also immobilized DNAzyme in a Au coated polycarbonate
track – etched (PCTE) nanocapillary array tube (NCAM) through thiolated enzyme.
In this case the cleaved fluorophore at the substrate strand has been analyzed in
solution from only as the cleaved moieties move on through the nanocapillary and
RESULTS AND DISCUSSION
131
hence follow aqueous state chemistry instead solid surface chemistry. Shen et al.
(2008) has developed an electrochemical biosensor for lead using DNAzyme and a
concentration of 1 µM 17 E was immobilized on the gold electrode and hybridized
with 2 µM 17 DS. In a recent study by Wang and Irudayaraj, (2011) DNAzyme based
lead biosensor has been constructed using surface enhanced raman scattering (SERS)
effect of gold nanopartical (AuNPs) in which enzyme strand was immobilized on gold
surface through thiolation followed by hybridization with AuNP labeled substrate
strand. The DNAzyme and raman reporters has been conjugated with AuNPs to
achieve a detection limit of 19.3 nM.
Finally the optimized enzyme and substrate strands were used to develop a
competent biosensor for lead detection. A linear range of detection was observed for
lead from 0.024 – 0.24 nanomoles (4.83 – 48.3 μM) as depicted in table 4.20 and
lower limit of detection achieved was 6 x 10-3
nanomoles (1.2 μM).
Table 4.20: Lead interaction with DNAzyme immobilized on gold surface after
17 E and 17 DS optimization
Lead Concentration
(nanomoles)
Fluorescence
(w.r.t acceptor against
HEX)
Change in fluorescence
(ΔF)
Control 0.777 ± 0.116
0.024 0.413 ± 0.129 0.364 ± 0.013
0.06 0.317 ± 0.112 0.460 ± 0.004
0.24 0.116 ± 0.134 0.661 ± 0.018
RESULTS AND DISCUSSION
132
As evident from the observations, the increase in lead concentration caused decrease
in fluorescence of acceptor as compared to the control. This was because the lead
catalyzed cleavage of the substrate strand which was acting as the donor in causing
FRET phenomenon and its cleavage resulted in break of FRET, also causing the
fluorescence of acceptor to decrease and change in fluorescence to increase. The
developed system was used to analyze lead in spiked milk samples and a detection
limit of 6 x 10-3
nanomoles (1.2 μM) was achieved in milk samples also (table 4.21),
but the change in fluorescence was low as compared to the synthetic system which
could be attributed to the complex matrix of milk which may interfere in fluorescence
signal.
Table 4.21: DNAzyme application on milk samples
Lead Conc.
(nanomoles)
Fluorescence Change in Fluorescence
(ΔF)
Control 1.705 ± 0.025
6 x 10-3
1.601 ± 0.038 0.104 ± 0.012
2.4 X 10-2
1.475 ± 0.022 0.230 ± 0.003
The advantage of using this DNAzyme based method for lead analysis in milk is that
there is no need of any pretreatment of milk to attain lead specificity and the complex
matrix of milk was still able to give detectable fluorescence signal unlike other optical
methods where removal of milk protein is a necessary step before analysis. DNAzyme
has been used to detect lead by many means either through gold nanoparticals (Liu
and Lu, 2003; Wei et al., 2008) or Fluorescent probes with or without quenchers (Liu
RESULTS AND DISCUSSION
133
and Lu, 2003; Chang et al., 2005). Single or both strands could be labeled with
fluorescent dyes to study the quenching or FRET phenomenon caused by Pb (II)
ions.
This DNAzyme approach provides high sensitivity and specificity towards Pb (II)
ions. DNAzyme has also been combined with network of microfluidic channels
coupled via a nanocapillary array to achieve a detection limit of 11 nM (Chang et al.,
2005). Wernette et al. (2005) had incorporated DNAzyme into PCTE-NCAMs and
obtained a detection limit of 16.9 nM. An electrochemical DNAzyme biosensor has
also been developed with detection limit of 0.3 µM (Xiao et al., 2006). DNAzyme has
also been entrapped in sol gel derived matrices, but detection limit was in micromolar
range (Shen et al., 2007). Conjugation of HRP mimicking DNAzyme with aptamers
has also been used for amplified Pb (II) detection, achieving a detection limit of 10
nM (Elbaz et al., 2008). Recent trend is the immobilization of DNAzyme on gold
surface to lower down the detection limit to 1 nM (Swearingen et al., 2005; Shen et
al., 2008). The present work has been focused to develop a micro-array based optical
DNAzyme lead biosensor for multiple sample analysis, in which the enzyme and
substrate strands has been labeled with fluorescent dyes and the FRET phenomenon in
the presence and absence of Pb (II) ions has been studied. Two Immobilization
strategies have been optimized for DNAzyme based studies. In first method
DNAzyme has been immobilized in TMOS based hydrosol gel in a microarray plate
and the microarray analysis of lead in hydrosol gel system has been successfully
miniaturized up to 10ul volume, with a detection limit of 0.24 nanomoles (24 μM).
This method has the advantage of multiple samples testing at the same time in a cost
effective way.
RESULTS AND DISCUSSION
134
Secondly, Immobilization of fluorophore bearing DNAzyme through
SAM on gold solid phase surface has been achieved. In this case, 6 x 10-3
nanomoles
(1.2 μM) biosensing of lead at 5ul volume has been achieved in synthetic samples and
milk samples. This method has an advantage of fast analysis along with testing of
several samples at the same time. The present biosensor has a novelty of sample
analysis directly on the gold chip as compared to the already developed samples in
which first the reaction is done on the gold surface and then transferred to other
system for analysis.
4.7.4 Mechanistic conformation of DNAzyme based lead analysis
The validation of Substrate strand cleavage was obtained by running the samples of
DNAzyme hybrid in the absence and presence of lead along with the individual 17 E
and 17 DS strands. As visible in the fig 4.30 the 33 bp 17 E strand (lane 2) was
obtained slightly above the 20 bp 17 DS strand (lane 1). Then the hybridized
DNAzyme incubated with deionized water (lane 4 & 8) showed a single band close to
17 E position. The samples incubated with lead (lane 5-7) showed the clear cleavage
of the substrate strand in which two distinct bands one corresponding to 17 E and the
lower one of cleaved 10 bp 17 DS were obtained. The cleaved 10 bp substrate strand
was obtained lower to the 20 bp 17DS strand confirming its cleavage in to short
fragments of 10 bp.
RESULTS AND DISCUSSION
135
Figure 4.30: PAGE analysis of 17 E and 17DS. Lane 1: 17S, Lane 2: 17E, Lane 4
& 8: Hybrid + water (control), lane 5-7: Hybrid + Pb2+
ions
4.8 Microarray based optical lead biosensor
To develop an optical microarray based biosensor for multiple sample analysis,
Rhodamine G (Ex = 526, Em = 555) was used as an indicator dye and read against
HEX (Ex = 535, Em = 556) in the plate reader. The dye was optimized for linear
concentration response on the microarray plate. A stock of 2mg/ 10 ml was diluted
340 times and then it’s different volume in a total of 680 µl was prepared and
emission read through plate reader. A volume of 80 µl dye in total of 680 µl was
optimized according to the linear curve (4.31) which showed the maximum emission
after which the intensity of fluorescence reached the pseudozero order stage.
RESULTS AND DISCUSSION
136
Figure 4.31: Calibration curve for Rhodamine-G dye
4.8.1 Analysis of lead in microarray plate
The lead was analyzed by setting the hydrosol- gel directly into the microarray plate
and then adding different lead concentrations in the wells with same initial
fluorescence. Whole cells of B. sphaericus was used as the sensing component with
enzyme activity 14.97 ± 2.01 IU (250 ml broth) and the fluorescent dye used was
Rhodamine – 6G (Ex = 526, Em = 555) which was read against HEX (Ex = 535, Em =
556) in the plate reader. The lead inhibition effect was found to be logarithmic in this
case also and the linear range of detection was found from 4.83 X 10-5
nanomoles to
4.8 X 10-2
nanomoles (4.83 nM – 4.83 μM ) as shown in figure 4.32.
The lower limit of detection in this method was 4.83 X 10-5
nanomoles (4.83 nM).
RESULTS AND DISCUSSION
137
Figure 4.32: Linear range of lead detection by direct hydrosol- gel
immobilization method
Although this method was sensitive but had the disadvantage of the selection of wells
with same fluorescence and results in many unutilised wells, hence wastage of
biomass and reagents. To overcome these limitations a modified method was
developed in which the hydrosol- gel was prepared in air tight tubes and lead was
added to them after all the required incubations. The analysis was done after pouring
20 µl from each reaction mixture in different wells of the microarray plate. This
method gave repeatability of the results and linear range of lead detection was
achieved upto to 4.8 X 10-4
to 9.7 X 10-3
nanomoles (0.48 – 9.66 nM), with lowest
detection limit of 4.8 X 10-4
nanomoles (0.48 nM) lead (figure 4.33).
Figure 4.33: Linear range detection of lead in modified microarray method
RESULTS AND DISCUSSION
138
Owing to the repeatibility and lower limit of detection of lead, this method was
further applied for lead analysis on milk samples (10 µl and 1000 µl ). In case of 1000
µl milk samples the linear range of detection was from 4.8 X 10-4
to 4.8 X 10-2
nanomoles (0.48 – 48.3 nM) in an logarithmic pattern as shown in figure 4.34 with
lowest limit of detection as 4.8 X 10-4
nanomoles (0.48 nM).
Figure 4.34: Linear range of detection of lead in spiked milk (1000 µl )
The linear range in 10 µl spiked milk was from 2.4 X 10-3
to 4.8 X 10-2
nanomoles
(2.4 – 48.3 nM) as shown in figure 4.35 and the lowest limit of detection obtained was
4.8 X 10-4
nanomoles (0.48 nM).
Figure 4.35: Linear range of detection of lead in spiked milk (10 µl)
RESULTS AND DISCUSSION
139
From the above shown data it is very evident that the developed method is suitable for
lead analysis in milk samples with a very low detection limit. The obtained limit of
detection in milk has not been reported by any other worker yet and the developed
method surpasses the need of any pretreatment to the milk samples. It was observed
that the complex matrix of milk imparts some increase in fluorescence of the dye as
compared to the synthetic system. So, to detect that interferrance due to milk and to
apply the developed method on unknown milk smples taking synthetic systen as
reference, the analysis of lead in synthetic and spiked milk samples was carried out
and the fluorescence read at different lead concentrations and in control samples was
compared. As shown in table 4.22, it was observed that the difference in fluoresence
of spiked milk and synthetic sample was following a regular pattern. So it could be
suggested that the effect of milk on fluorescence is corresponding to 0.12 which
should be considered at the time of lead determination in unknown milk samples.
Table 4.22: Comparison of fluorescence in synthetic and spiked milk samples
Lead
Concentration
(nanomoles)
Fluorescence in
Synthetic system
Fluorescence in
spiked milk
sample (10 μl)
Fluorescence
difference
Control 0.502 0.641 0.139
4.8 X 10-4
0.525 0.654 0.129
2.4 X 10-3
0.534 0.653 0.119
4.8 X 10-3
0.547 0.667 0.120
9.7 X 10-3
0.553 0.671 0.119
2.4 X 10-2
0.581 0.683 0.102
Av=0.12
RESULTS AND DISCUSSION
140
The detection limit achieved in the developed biosensor is far below the limits
obtained till now and is under the permissible limit of lead in milk (96.6 nM or 20 ng
ml-1
). No optical based study has been done so far for lead contamination in milk. So
the comparison has been done with studies on water samples. Lee and Lee, 2002 had
developed a conductometric biosensor based on sol- gel immobilized urease and
obtained a detection limit of 0.9 mM for lead. Later Tsai et al. (2003) used FITC-
dextran as fluorescent dye to develop urease based optical biosensor for heavy metals
and achieved a detection limit of 0.1 mM. Kuswandi (2003) obtained a detection limit
of 1 X 10-2
mM lead in 6 min response time using immobilized urease on a optical
fiber. Then a sol- gel immobilized urease conductometric biosensor was developed by
Illangovan et al. (2006) and a percentage inhibition of 35% was observed with 1 mM
lead. A very low detection limit of 4.83 nM (4.83 X 10-3
nanomoles) was achieved
using cyclotetrachronotropylene (CTCT) as an indicator by Haron and Ray (2006).
They developed an optical biosensor based on total reflection at interface between
SiN4 core and composite polyelectrolyte self assembled (PESA) membrane caused by
the catalytic activities of urease and acetylcholine esterase.
4.8.2 Detection of lead in the presence of cadmium
The detection of lead and cadmium simultaneously was possible by using two urease
producing strains, B. sphaericus and B. badius respectively. Two different fluorescent
dyes Rhodamine 6G (Ex = 526, Em = 555) and Acridine orange (Ex = 493, Em = 535)
with different emission wavelengths were selected for this purpose and were read
against HEX (Ex = 535, Em = 556) and Sybergreen (Ex = 497, Em = 520 nm)
respectively. Also there was an advantage that B. badius urease is insensitive to lead
at nanomolar levels and only shows inhibition at micromolar range, wheres as B.
sphaericus is sensitive to lead at nanomolar levels. So lead does not interfere in
RESULTS AND DISCUSSION
141
cadmium detection by B. badius and the fluorescence read against sybergreen in the
system corresponds to cadmium effect only, but the fluorescence read against HEX is
the cumulative effect of lead and cadmium on B. sphaericus. The standard curve of
cadmium was prepared for synthetic and spiked milk samples (figure 4.36 and figure
4.37).
Figure 4.36: Cadmium standard curve in simultaneous study with Pb (II) in
synthetic system
Figure 4.37: Cadmium standard curve in simultaneous study with Pb (II) in
spiked milk samples (10 µl)
RESULTS AND DISCUSSION
142
The lead concentration in an unknown sample was detected by carrying out
simultaneous detection of lead and cadmium in the synthetic and spiked milk samples.
According to the standard curve of cadmium effect on urease, its concentration could
be deduced from the change in fluorescence pattern against sybergreen dye
(corresponding to Cd2+
). Taking that concentration as reference the respective
fluorescence fraction of Cd2+
was subtracted against the HEX fluorescence to get the
lead contamination value. The present method is the first of its kind for simultaneous
detection of Pb (II) and Cd (II) in milk in the same well. Haron and Ray, (2006) had
developed an optical biosensor for monitoring Pb (II) and Cd (II) in water samples
based on inhibition of urease and acetylcholinesterase and achieved a detection limit
of 1ng ml-1
whereas the present work could detect upto 0.1 ng ml-1
in milk samples.
4.8.3 Application of developed sample on different areas milk samples
Twenty milk samples each from rural, urban and industrial area were collected and
lead analysis was done on them according to the developed method. As per the
standards of spiked lead samples and cadmium interference values, two formulas for
quantitative analysis of lead was devised which could be directly applied on unknown
samples.
X nanomoles of Pb2+
= [ (A – M1) – F1] X B1 , when only Rhodamine dye used
Where, A = Fluorescence read for the sample (corresponding to Pb2+
)
M1 = Milk factor with Rhodamine dye,
F1 = Fluorescence read of Rhodamine G
B1 = Pb2+
Standard / Fluorescence of that standard sample
To detect lead in the presence of cadmium a formula was devised for cadmium
detection and then the fluorescence factor (interference factor) according to that
RESULTS AND DISCUSSION
143
cadmium concentration was subtracted from Rhodamine G fluorescence to detect lead
contamination.
Y nanomoles of Cd2+
= [(C – M2) – F2] X B2
Where, C = Fluorescence read for the sample (corresponding to Cd2+
),
M2 = Milk factor with acridine orange dye,
F2 = Fluorescence read of acridine orange dye
B2 = Cd2+
Standard / Fluorescence of that standard sample
Formula for detection of lead in the presence of cadmium
X nanomoles of Pb2+
= [ {[ A – M1) – F3] – C} X B1
Where, A = Fluorescence read for the sample (corresponding to Pb2+
),
M1 = Milk factor,
F3 = Florescence read of combined dyes rhodamine and acridine orange
C = Cadmium conc. calculated from standard curve using dye acridine orange,
B1 = Pb2+
Standard / Fluorescence of that standard sample
The reliability of the developed formula was checked by carrying out analysis of lead
and cadmium in unknown samples and then again after spiking of the samples with
known concentration of lead and cadmium. The lead analysis was done using
Rhodamine G dye alone and in the presence of acridine orange to detect the presence
of cadmium simultaneously in the milk samples. All the samples were analyzed in
triplicates and the devised formula was applied to the fluorescence read to detect lead
and cadmium contamination.
All the rural and urban samples were found free of any lead or cadmium
contamination (table 4.23 and table 4.24). In the case of industrial samples lead and
cadmium was found in two samples at the levels shown in the table 4.25.
RESULTS AND DISCUSSION
144
Table 4.23 Lead and cadmium contamination in rural samples
Sample Place Lead detected
(nanomoles)
Lead detected
(nanomoles) in
the presence of
Cd2+
Cadmium
detected
(nanomoles)
Sample R1 VPO- Banur,
Rajpura
ND ND ND
Sample R2 VPO- Banur,
Rajpura
ND ND ND
Sample R3 VPO- Banur,
Rajpura
ND ND ND
Sample R4 VPO- Banur,
Rajpura
ND ND ND
Sample R5 VPO- Banur,
Rajpura
ND ND ND
Sample R6 VPO- Banur,
Rajpura
ND ND ND
Sample R7 VPO- Banur,
Rajpura
ND ND ND
Sample R8 VPO- Banur,
Rajpura
ND ND ND
Sample R9 VPO- Banur,
Rajpura
ND ND ND
Sample R10 VPO- Banur,
Rajpura
ND ND ND
Sample R11 VPO- Banur,
Rajpura
ND ND ND
Sample R12 VPO- Banur,
Rajpura
ND ND ND
Sample R13 VPO- Banur,
Rajpura
ND ND ND
Sample R14 VPO- Banur,
Rajpura
ND ND ND
Sample R15 VPO- Banur,
Rajpura
ND ND ND
Sample R16 VPO- Banur,
Rajpura
ND ND ND
Sample R17 VPO- Banur,
Rajpura
ND ND ND
Sample R18 VPO- Banur,
Rajpura
ND ND ND
Sample R19 VPO- Banur,
Rajpura
ND ND ND
Sample R20 VPO- Banur,
Rajpura
ND ND ND
*ND = No heavy metal detected
RESULTS AND DISCUSSION
145
Table 4.24 Lead and cadmium contamination in urban samples
Sample Place Lead
detecte
d
(micro
moles)
Lead detected
(micromoles)
in the
presence of
Cd2+
Cadmium
detected
(nanomoles)
Sample U1 Tripuri, Patiala ND ND ND
Sample U2 Tripuri, Patiala ND ND ND
Sample U3 Tripuri, Patiala ND ND ND
Sample U4 Tripuri, Patiala ND ND ND
Sample U5 Tripuri, Patiala ND ND ND
Sample U6 Tripuri, Patiala ND ND ND
Sample U7 Tripuri, Patiala ND ND ND
Sample U8 Tripuri, Patiala ND ND ND
Sample U9 Tripuri, Patiala ND ND ND
Sample U10 Tripuri, Patiala ND ND ND
Sample U11 Urban State Ph- II,
Patiala
ND ND ND
Sample U12 Urban State Ph- II,
Patiala
ND ND ND
Sample U13 Urban State Ph- II,
Patiala
ND ND ND
Sample U14 Urban State Ph- II,
Patiala
ND ND ND
Sample U15 Urban State Ph- II,
Patiala
ND ND ND
Sample U16 Urban State Ph- II,
Patiala
ND ND ND
Sample U17 Urban State Ph- II,
Patiala
ND ND ND
Sample U18 Urban State Ph- II,
Patiala
ND ND ND
Sample U19 Urban State Ph- II,
Patiala
ND ND ND
Sample U20 Urban State Ph- II,
Patiala
ND ND ND
RESULTS AND DISCUSSION
146
Table 4.25 Lead and cadmium contamination in industrial samples
Sample Place Lead
detected
(nanomoles)
Lead detected
(nanomoles) in
the presence of
Cd2+
Cadmium
detected
(nanomoles)
Sample D1 Industrial Area,
Ludhiana
ND ND ND
Sample D2 Industrial Area,
Ludhiana
ND ND ND
Sample D3 Industrial Area,
Ludhiana
ND ND ND
Sample D4 Industrial Area,
Ludhiana
ND ND ND
Sample D5 Industrial Area,
Ludhiana
ND ND ND
Sample D6 Industrial Area,
Ludhiana
5.9 ± 0.32 7.1 ± 0.29 0.026 ± 0.022
Sample D7 Industrial Area,
Ludhiana
6.0 ± 0.33 6.2 ± 0.25 ND
Sample D8 Industrial Area,
Ludhiana
ND ND 0.162 ± 0.039
Sample D9 Industrial Area,
Ludhiana
ND ND ND
Sample D10 Industrial Area,
Ludhiana
ND ND ND
Sample D11 Tajpur Road,
Ludhiana
ND ND ND
Sample D12 Tajpur Road,
Ludhiana
ND ND ND
Sample D13 Tajpur Road,
Ludhiana
ND ND ND
Sample D14 Tajpur Road,
Ludhiana
ND ND ND
Sample D15 Tajpur Road,
Ludhiana
ND ND ND
Sample D16 Tajpur Road,
Ludhiana
ND ND ND
Sample D17 Tajpur Road,
Ludhiana
ND ND ND
Sample D18 Tajpur Road,
Ludhiana
ND ND ND
Sample D19 Tajpur Road,
Ludhiana
ND ND ND
Sample D20 Tajpur Road,
Ludhiana
ND ND ND
RESULTS AND DISCUSSION
147
Reliability of the developed method and formulas was checked by applying the same
on three unknown samples D4, D19, D20 (which were detected negative for both Pb2+
and Cd2+
earlier) after spiking with 5 ng lead and cadmium each. The results are
shown in table 4.26.
Table 4.26: Reliability studies of Optical Biosensor
Sample Pb2+
and
Cd2+
Added
(ng)
Lead found
(ng)
Cadmium found (ng)
Sample D4 5 4.64 ± 0.33 4.69 ± 0.38
Sample D19 5 4.88 ± 0.26 5.01 ± 0.17
Sample D20 5 5.3 ± 0.10 4.77 ± 0.19
Finally the present work has developed new technologies for the successful
construction of enzyme inhibition based colorimetric, potentiometric, amperometric
and optical biosensor along with DNAzyme based optical biosensor for monitoring
lead in milk.