SondeCAM User Guide
Transcript
Page 1: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

SondeCAM User Guide

Page 2: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

2091 Exchange Court i Fairborn, Ohio 54324

937-426-2703 www.FishSens.com

Table of Contents Overview ......................................................................................................................................... 1

SondeCAM mini ........................................................................................................................... 1

SondeCAM ................................................................................................................................... 1

SondeCAM WQ ............................................................................................................................ 1

Boat Installation .............................................................................................................................. 2

Generic Configuration ................................................................................................................. 2

SondeCAM WQ Lowrance® Configuration .................................................................................. 3

Using SondeCAM Accessories ......................................................................................................... 4

SondeCAM DVR ........................................................................................................................... 4

Topside Rail Mount Kit ................................................................................................................ 6

Boatside Rail Mount Kit ............................................................................................................... 6

Pole Mount Kit............................................................................................................................. 6

Panel Mount Plate ....................................................................................................................... 6

Flashlight Kit ................................................................................................................................ 6

Sinking Weight Kit ....................................................................................................................... 6

Trolling Motor Mount Kit ............................................................................................................ 6

SondeCAM WQ Setup and Operation ............................................................................................. 7

Selecting a Water Quality Parameter.......................................................................................... 7

Calibration ................................................................................................................................... 7

SondeCAM WQ Fishfinder/Chartplotter Setup .............................................................................. 9

Lowrance® ................................................................................................................................... 9

Maintenance ................................................................................................................................. 11

Camera ...................................................................................................................................... 11

Water Quality Sensors ............................................................................................................... 12

Troubleshooting ............................................................................................................................ 16

Generic ...................................................................................................................................... 16

SondeCAM WQ .......................................................................................................................... 16

Warranty and Service.................................................................................................................... 17

Overview ................................................................................................................................... 17

Page 3: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

2091 Exchange Court ii Fairborn, Ohio 54324

937-426-2703 www.FishSens.com

Limitation of Warranty .............................................................................................................. 17

Warning ..................................................................................................................................... 17

Authorized U.S. Service Centers ................................................................................................ 17

Mechanical Drawings and Specifications ...................................................................................... 18

SondeCAM mini ......................................................................................................................... 18

SondeCAM ................................................................................................................................. 19

SondeCAM WQ .......................................................................................................................... 20

Appendix ....................................................................................................................................... 22

SondeCAM WQ Direct Communication .................................................................................... 22

Revision History ......................................................................................................................... 23

Page 4: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

2091 Exchange Court iii Fairborn, Ohio 54324

937-426-2703 www.FishSens.com

PAGE INTENTIONALLY LEFT BLANK

Page 5: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Overview SondeCAM mini

1

Overview

SondeCAM mini SondeCAM mini is a small profile and affordable option for underwater video with direct connection to on-board electronics.

SondeCAM SondeCAM is the original FishSens underwater video camera with direct connection to on-board electronics.

SondeCAM WQ SondeCAM WQ incorporates video and water quality measurements with direct connection to on-board electronics.

Page 6: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Boat Installation Generic Configuration

2

Boat Installation

Generic Configuration

1 SondeCAM & mini 6 5.5 X 2.1 mm Female Barrel Connector

2 Fishfinder (Not included)

7 6 ft. Power Cable

3 12V Battery (Not included)

8 Fuse, Glass, 1A, 250V, 5x20 mm, Slow-Blow

Diode, Schottky, 1A, 20V

4 Power & Video Adapter 9 RCA Male to RCA Male Adapter (Optional)

5 RCA Female 10 RCA Male to BNC Male Adapter (Optional)

Connect the SondeCAM system as show above.

Power the Fishfinder and open the video streaming interface.

Page 7: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Boat Installation SondeCAM WQ Lowrance® Configuration

3

SondeCAM WQ Lowrance® Configuration

1 SondeCAM 7 Butt Connector

2 Lowrance® 8 Lowrance® Fuse

3 12V Battery 9 3A PTC Resettable Fuse

4 Lowrance® Video Cable 10 UW6P Connector

5 Lowrance® Power Cable 11 RCA Female

6 FishSens Lowrance® Adapter 12 RCA Male

Connect the SondeCAM system as show above.

Power the Lowrance® and select Video from the main menu.

For SondeCAM WQ applications, seeSondeCAM WQ Fishfinder/Chartplotter Setup:

Lowrance®.

Page 8: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Using SondeCAM Accessories SondeCAM DVR

4

Using SondeCAM Accessories

SondeCAM DVR

Connecting the DVR

1. Unscrew the SD card port plug.

2. Insert a SD Card and reinstall the plug.

3. Connect the SondeCAM to the DVR receptacle.

4. Connect the DVR plug to the fishfinder adapter.

Record Button

Press the Record Button to begin a recording.

o A red circle will blink in the top left corner of the display

while recording.

Press the Record Button again to stop recording.

DVR Operation

Use the remote to move through the menu and make selections.

RECODE: Start/stop recording OR select

MODE: Enter/exit playback mode

MENU: Open/close menu

UP: Move curser up

DOWN: Move curser down

Page 9: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Using SondeCAM Accessories SondeCAM DVR

5

DVR Menu

To enter the menu click MENU.

Resolution

Choose the video resolution (influences video output quality).

D1: 720x480

VGA: 640x480

QVGA: 320x240

Enable/Disable Date Stamp

Choose whether the date and time will be shown at the bottom of the

recorded video.

Image Quality

Allows the user to select the quality of the recorded video.

Influences size of recorded video files.

DVR Advanced Menu

To enter the Advanced Menu click MENU twice from the live video

screen.

Format SD Card

Erases all data on the SD Card and formats it for use with the

SondeCAM DVR.

Set Date/Time

Use UP and DOWN to change.

Use REC/OK to move to next.

Use MENU to exit.

Page 10: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Using SondeCAM Accessories Topside Rail Mount Kit

6

Topside Rail Mount Kit 1. Attach the RAM ball (b) to the camera with the supplied lock washer.

2. Mount the u-bolt assembly I to the boat rail. (Fits 0.75” to 1.25” diameter

rails).

3. Attach the SondeCAM (a) to the u-bolt assembly (c) with the (d).

Boatside Rail Mount Kit 1. Attach the RAM ball (b) to the camera with the supplied lock washer.

2. Mount the u-bolt assembly I to the boat rail. (Fits 0.75” to 1.25”

diameter rails).

3. Attach the SondeCAM (a) to the mast I with single socket arm (f).

4. Attach the mast I to the u-bolt assembly I with the double socket arm (d).

Pole Mount Kit 1. Attach the RAM ball (b) to the camera with the supplied lock washer.

2. Connect the threaded adapter (g) to the telescoping handle (h).

3. Attach the Tele-Mount (i) to the threaded adapter (g).

4. Attach the Tele-Mount (i) and SondeCAM (a) with the double socket arm (d).

Panel Mount Plate 1. Using the plate (j) as a template, drill a 1.5” diameter hole and (4) corner

holes for ¼” screws.

2. Mount the plate (j) to the console with the supplied ¼” screws.

3. Feed the Interface Cable (k) through the plate (j) and secure with the

supplied nut (l).

Flashlight Kit 1. Attach the flashlight mount top (m) to the SondeCAM (a) with the

supplied ¼” screw.

2. Orient Pelican flashlight (n) so it fits into the mount and secure the

flashlight mount bottom (o) with supplied #6 screws.

Sinking Weight Kit 1. Attach the weight (p) to the bottom of the SondeCAM (a) with the supplied

¼” screw.

Trolling Motor Mount Kit 1. Attach the RAM ball (b) to the camera with the supplied lock washer.

2. Loosely clamp the mount (q) to the RAM ball using the screws and insert

nuts I.

3. Strap the mount assembly to the trolling motor with the hose clamp (s).

4. Adjust the camera position and angle and tighten the clamp securely to

the RAM ball.

5. Zip tie the camera cable.

Page 11: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter

7

SondeCAM WQ Setup and Operation

Selecting a Water Quality Parameter 1. Remove the probe guard and port plug from the

bulkhead.

2. Install the pH or dissolved oxygen probe.

3. Insert the magnet (back of screwdriver) into the

bulkhead hole and observe the LED color.

Red = pH

Green = DO

Hold magnet in bulkhead hole for 30 seconds

until LED color changes to switch modes.

Page 12: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

SondeCAM WQ Setup and Operation Calibration

8

Calibration

Warning: The pH and DO sensors used with the SondeCAM WQ are precise scientific

instruments and care must be taken during calibration in order to ensure accurate readings.

The integrated precision temperature sensor does not require calibration. The depth sensor is

calibrated automatically during either pH or dissolved oxygen calibration.

pH Calibration

1. Place the pH probe in pH calibration solution.

Can be done with pH 4, 7, and/or 10 solution.

Starting with pH 7 is recommended.

2. Hold magnet in bulkhead hole for 3 seconds until LED

begins to blink red.

3. Remove magnet.

4. Wait for the LED to stop blinking red (about 45 seconds).

There is an automatic retry if the first attempt fails.

If the retry fails, the red LED will remain on.

5. Repeat for other calibration solutions.

Any un-calibrated points will remain at ideal values.

Note: When handling pH electrodes, rinse the electrodes with distilled water before and after

each calibration standard. Blot the end of the electrode with lint-free cloth to remove excess

water. Never wipe the electrode to remove excess water – wiping can create static charges that

interfere with correct pH measurement.

DO Calibration

1. Expose DO probe to 100% air environment

A bag with a wet sponge or paper towel

works well.

In open air the DO value changes frequently

and calibration may fail.

2. Hold magnet in bulkhead hole for 3 seconds until

LED begins to blink green.

3. Wait for the LED to stop blinking green (about 45 seconds).

There is an automatic retry if the first attempt fails.

If the retry fails, the green LED will remain on.

Page 13: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

SondeCAM WQ Fishfinder/Chartplotter Setup Lowrance®

9

SondeCAM WQ Fishfinder/Chartplotter Setup

Lowrance® 1. From the start screen, click the Video button to view

live streaming video.

2. Swipe down the blue menu button in the top

right to open the menu.

3. Click Data Overlay if not already selected.

4. Click Edit Overlay.

5. Click Add….

6. Expand GPS and check

Altitude.

7. Expand Sonar and check

Water Temperature and

Depth.

8. Expand Weather and check

Wind Speed.

9. Click the dropdown next to

one of the expanded items

and click Data Sources….

Page 14: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

SondeCAM WQ Fishfinder/Chartplotter Setup Lowrance®

10

10. Expand Sonar, GPS, and

Weather.

11. Set Altitude, Water

Temperature, Depth, and

Wind Speed to HDS-9T

NMEA0183 Port 1 [This

device].

12. Click Close.

13. Click Save.

14. Set the correct units for the

selected parameters in

Settings | Units.

Temperature can be

set to °C or °F

Depth can be set to

Meters or Feet

Wind Speed MUST

be set to Knots

Altitude MUST be set

to Meters

15. Return to Video to view live

data and video.

Data Mapping

DO / pH data is mapped to Altitude and Wind Speed as shown below.

Altitude Wind Speed

pH mV pH Units

DO mg/L %

Page 15: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Maintenance Camera

11

Maintenance

Camera Rinse the device in clean water and wipe gently with non-abrasive cloth after use to

remove any buildup

Wipe the lens with a non-abrasive cloth if the image becomes foggy

Ensure all connections are dry and secure

Re-apply O-ring grease to underwater 6-pin plug connectors periodically

Page 16: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Maintenance Water Quality Sensors

12

Water Quality Sensors

DO Probe (YSI 2002 Galvanic Dissolved Oxygen Sensor)

Preparing for Measurement

1. Remove the probe guard to access the sensor and unscrew from the SondeCAM.

2. Unscrew and discard any old Membrane Cap at the bottom of the electrode.

3. Rinse the internal anode/cathode element with distilled water, blot dry.

4. Fill a new Membrane Cartridge with the provided DO Electrolyte solution.

5. Fill the new Membrane Cap with DO electrolyte and gently tap the edge to release any

trapped air bubbles.

6. Finger-tighten until snug. Do not over tighten.

7. Rinse the assembled electrode with distilled water.

8. Immerse the tip of the electrode in de-ionized water and stir the water for 5 minutes to

properly clean the electrode.

Note: When preparing the sensor, be careful not to touch the membrane with your fingers. The

oils on your skin can affect the dissolution of oxygen through the membrane and in turn affect

the readings of the sensor.

Sensor Storage

Short Term (1-3 Days)

The assembled electrode should be stored in a moist environment, such as in the

provided calibration/storage bottle cup or in a beaker of water. When storing in

calibration/storage cup, place a small piece of sponge inside. Moisten sponge with

distilled or de-ionized water.

Long Term

Completely disassemble the electrode. Rinse the anode/ cathode element and

Membrane Cap with distilled or de-ionized water. Blot dry all of the parts. The

Membrane Cap may be placed on the electrode, but do not tighten. Store all parts

securely in the original box.

When ready to use the probe again, follow the instructions in the “Preparing DO Sensors

for Measurement” section.

Note: Galvanic DO sensors continually reduce oxygen regardless of if the SondeCAM is

powered. This allows for the sensor to report instantaneous measurements when the

SondeCAM receives power. As a result, the sensor continues to oxidize even when not in use.

The oxidation will result in solids forming and the solution appearing milky white. This is normal

and will not interfere with the sensor’s reading unless the buildup becomes excessive.

Page 17: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Maintenance Water Quality Sensors

13

Cleaning and Reconditioning

Over time, deposits will develop on the silver anode and gold cathode due to the reduction of

oxygen. These deposits are easily removed with 400 grit wet/dry sand paper.

1. Remove the probe guard to access the sensor and unscrew from the SondeCAM.

2. Unscrew and discard the old Membrane Cap at the bottom of the electrode.

3. Rinse the internal anode/cathode element with distilled water, blot dry.

4. Use 400 grit wet/dry sand paper to polish the gold cathode and silver anode.

Gold cathode: Place dampened sand paper face up in your palm. Holding the sensor

vertically, bring the gold cathode tip straight down onto the sand paper and twist. The

purpose is to remove build-ups and lightly scratch the cathode. The gold should have a

matte finish when complete.

Silver anode: Place dampened sand paper on tip of sensor, apply light finger pressure,

and twist in a circular motion (usually 3-4 twists is sufficient). The purpose is to remove

the build-ups without scratching or damaging the anode.

5. After polishing, thoroughly rinse with distilled or de-ionized water and blot dry.

6. Fill the new Membrane Cap with DO electrolyte and gently tap the edge to release any

trapped air bubbles.

7. Finger-tighten until snug. Do not over tighten.

8. Rinse the assembled electrode with distilled water.

Note: When cleaning the sensor, be careful not to touch the membrane with your fingers. The

oils on your skin can affect the dissolution of oxygen through the membrane and in turn affect

the readings of the sensor.

If the above procedure does not work, please contact FishSens Technology for assistance.

Page 18: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Maintenance Water Quality Sensors

14

pH Probe (YSI 1001A Amplified pH Sensor)

Preparing for Measurement

pH sensors are shipped with the pH bulb moist. Prior to using your WQ-pH for the first time,

remove the pH storage bottle from the bottom of the electrode and rinse the electrode with

distilled or de-ionized water (Note: Keep the pH storage bottle for late storage). The electrode

is now ready for calibration.

Note: When handling pH electrodes, rinse the electrodes with distilled water before and after

measuring a sample. Blot the end of the electrode with lint-free cloth to remove excess water.

Never wipe the electrode to remove excess water – wiping can create static charges that

interfere with correct pH measurement.

Sensor Storage

Short Term Storage

Between measurements, store the pH electrode in the pH storage bottle containing pH

4.00 buffer.

Long Term Storage

When storing for longer periods, store the pH electrode in the pH storage bottle which

came with the electrode. The storage bottle should contain pH 4 buffer.

Do not store the electrode in distilled or deionized water – this will cause ions to leach

out of the glass bulb and render your sensor useless. After storage, you may notice

white KCl crystals deposited on your electrode. Such salt formation will not interfere

with measurements. Simply rinse the electrode with distilled water to remove the

crystals and blot dry before use.

Cleaning and Reconditioning

As the pH sensor ages, it may exhibit sluggish or noisy readings. Cleaning is recommended

when deposits form on the glass and/or platinum surfaces. Try the below procedures in order

and stop once good readings are achieved.

Wipe Clean

Use a moistened cloth/lens cleaning tissue/cotton swab to remove any contaminants

from the probe. Do NOT apply pressure to the glass bulb as it may break.

Wash in Soapy Water

1. Let the probe sit in a solution of liquid soap, (about 1⁄2 teaspoon per 200mL warm

water).

Page 19: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Maintenance Water Quality Sensors

15

2. Using a soft cloth to gently wipe the pH glass – remember that pH glass is extremely

delicate and breaks very easily.

3. Rinse the sensor in clean water when done.

Soak in Hydrochloric Acid

1. Place the probe in one molar (1 M) hydrochloric acid (HCL).

2. Let sit for 30-60 minutes.

3. Rinse in clean water (not DI).

4. Wipe clean with a cotton swab.

5. Re-rinse in clean water.

6. Soak in clean water for an hour. Stir periodically.

Soak the WQ-pH sensor in a beaker of warm water (50°C) for 15 minutes to remove

dried gel or salts from the junction. Then place in a beaker of warm 4M KCl solution. Set

aside until it returns to room temperature. The gel should be moist and the junction

flow should be restored.

Soak in Chlorine Bleach

WARNING: Do not mix the HCL with the chlorine bleach. A toxic gas product can form.

Rinse thoroughly before proceeding.

1. Place the probe in 1:1 dilution of commercial chlorine bleach.

2. Let sit for 60 minutes.

3. Rinse in clean water (not DI).

4. Soak in clean water for 60 minutes or more to remove all traces of bleach. Stir

periodically.

5. Re-rinse in clean water.

If the above procedure does not work, please contact FishSens Technology for assistance.

Page 20: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Troubleshooting Generic

16

Troubleshooting

Generic Issue Steps to Resolve

Video does not appear on the display

1. Confirm there is nothing blocking the camera lens 2. Confirm all connections are secure 3. Confirm the boat battery is good 4. Connect the camera’s RCA output to another display’s (Such as

a TV) RCA yellow input to confirm the camera is the issue

Camera body feels warm

This is normal with prolonged use in air and is not an issue. However, the body can be cooled through the following methods:

1. Remove power and let the unit cool down 2. Place the unit in water

SondeCAM WQ Issue Steps to Resolve

Water quality values are unreasonable

1. Confirm correct mode of operation (Red LED = pH; Green LED = DO)

2. Calibrate the sensor 3. Check the sensor for damage 4. Re-condition

Page 21: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Warranty and Service Overview

17

Warranty and Service

Overview FishSens Technology, Inc. warrants the instruments it manufactures against defects in materials

or workmanship for a period of 1 year from the date of delivery to the original customer. This

warranty is limited to the replacement or repair of such defects, without charge, when the

instrument is returned to FishSens Technology, Inc. Damage due to accidents, misuse,

tampering, lack of reasonable care, loss of parts, failure to perform prescribed maintenance, or

acts of nature is not covered. This warranty excludes all other warranties, express or implied,

and is limited to a value not exceeding the purchase price of the instrument.

Limitation of Warranty This warranty is not applicable to any FishSens Technology, Inc. product damage or failure

caused by (i) failure to install, operate or use the product in accordance with FishSens

Technology, Inc. written instructions, (ii) abuse or misuse of the product, (iii) failure to maintain

the product in accordance with FishSens Technology, Inc. written instructions, (iv) any improper

repairs to the product, (v) use by you of defective or improper components or parts in servicing

or repairing the product, or (vi) modification of the product in any way not expressly authorized

by FishSens Technology, Inc.

Warning FishSens Technology, Inc. products are not authorized for use as critical components in any life

support system where failure of the product is likely to affect its safety or effectiveness.

Authorized U.S. Service Centers FishSens Technology, Inc.

2091 Exchange Court, Fairborn, Ohio 45324

Phone: (937) 426-2151

Fax: (937) 426-1125

E-Mail: [email protected]

Page 22: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Mechanical Drawings and Specifications SondeCAM mini

18

Mechanical Drawings and Specifications

SondeCAM mini Mechanical Drawing

Specifications

Weight 1.0 lbs with cable

Cable Length 25 ft. included, 25 ft., 50 ft., 100 ft.

extension cables available

Buoyancy Sinks without additional weights

Material Marine anodized aluminum, acetal,

stainless steel hardware

Camera Single Chip 1/3” CMOS

Resolution >500 TV Line

Noise Reduction 2D + 3D – NR

Shutter 1/50 to 1/100000 s

Minimum Color Illumination 0.1 lux

Minimum B/W Illumination 0.0008 lux

Video Output NTSC/PAL

Voltage 12VDC ± 10%

Power 2.0 Watts

Max Depth 300 ft.

Page 23: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Mechanical Drawings and Specifications SondeCAM

19

SondeCAM Mechanical Drawing

Specifications

Weight 2.8 lbs with cable

Cable Length 50 ft. included, 25 ft., 50 ft., 100 ft.

extension cables available

Buoyancy Sinks without additional weights

Material Marine anodized aluminum, acetal,

stainless steel hardware

Camera 1/3” CCD Sony Effio-V CXD4141GG DSP

Resolution >700 TV Line

Noise Reduction 2D + 3D – NR

Shutter 1/10000 s

Minimum Color Illumination 0.1 lux

Minimum B/W Illumination 0.001 lux

Video Output NTSC/PAL

Voltage 12VDC ± 10%

Power 2.0 Watts

Max Depth 300 ft.

Page 24: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Mechanical Drawings and Specifications SondeCAM WQ

20

SondeCAM WQ Mechanical Drawing

Specifications

Weight 2.8 lbs with cable

Cable Length 50 ft. included, 25 ft., 50 ft., 100 ft.

extension cables available

Buoyancy Sinks without additional weights

Material Marine anodized aluminum, acetal,

stainless steel hardware

Camera 1/3” CCD Sony Effio-V CXD4141GG DSP

Resolution >700 TV Line

Noise Reduction 2D + 3D – NR

Shutter 1/10000 s

Minimum Color Illumination 0.1 lux

Minimum B/W Illumination 0.001 lux

Video Output NTSC/PAL

Voltage 12VDC ± 10%

Power 2.0 Watts

Max Depth 300 ft.

Operating Temperature 0 to 50°C (32 to 122°F)

Page 25: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Mechanical Drawings and Specifications SondeCAM WQ

21

Depth Sensor Specifications

Depth Sensor MS5803-05BA Miniature Altimeter and

Diving Module

Range 0 to 6 bar

Resolution 0.1ft

Operating Temperature 0 to 50°C (32 to 122°F)

Response 8.22 ms

Accuracy ± 2% of the reading (±120 mbar)

Temperature Sensor Specifications

Temperature Sensor Thermistor

Range 0 to 45°C (32 to 113°F)

Resolution 0.1°

T90 Response Time 60 sec

Accuracy ±0.075°C

Dissolved Oxygen Sensor Specifications (optional)

Dissolved Oxygen Sensor YSI 2002 Galvanic Dissolved Oxygen

Sensor

pH Sensor Specifications (optional)

pH Sensor YSI 1001a Amplified pH

Page 26: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Appendix SondeCAM WQ Direct Communication

22

Appendix

SondeCAM WQ Direct Communication Advanced setup and interfacing is possible using a RS-485 adapter cable and a terminal

program. Data beings streaming automatically when power is applied. Advanced setup can be

accessed by typing MENU in a terminal interface. Within 2 seconds a menu will appear

providing setup and configuration options.

NOTE: This menu provides access to the device’s root programming. FishSens Technology, Inc.

is not responsible for any permanent damage caused by incorrectly changing the setup.

Please contact technical support ([email protected]) for assistance.

Hardware Protocol: RS485 N-8-1 at 4800 baud

Software Protocol: NMEA 0183

o Message MTW: Temperature (°C)

o Message DPT: Depth (meters)

o Message MWV: DO OR pH (% OR pH Unit)

o Message GGA: DO OR pH (mg/L OR pH mV)

Cable Pinout

UW6 Plug UW6 Receptacle

1 NC - 1 RS485A (Non-Inverting)

Green

2 Video Blue 2 RS485B (Inverting)

Green Stripe

3 GND/Video Shield/Shield

Brown/Brown Stripe/ Blue Stripe/Bare

3 12V+ Yellow/Yellow Stripe

4 12V+ Yellow/Yellow Stripe 4 GND/Video Shield

Brown/Brown Stripe/ Blue Stripe

5 RS485B (Inverting)

Green Stripe 5 Video Blue

6 RS485A (Non-Inverting)

Green 6 NC -

Connectors are drawn looking into the connector

Single circles are pins and double circles are sockets

Page 27: SondeCAM User Guide - FishSens TechnologySondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 7 SondeCAM WQ Setup and Operation Selecting a Water Quality Parameter 1.

Appendix Revision History

23

Revision History Revision Date Code Changes Reviser

0.1 15B03 Original Document Created KS

1.0 15B12 Updated WQ Maintenance Section KS

1.1 15C02 Updated Wiring Diagram KS

1.2 15C20 Updated Wiring Diagram, DVR Section, Dimensions, Cable Pinout, General Clarity

KS

1.3 15C31 Updated Accessories, SC mini weight KS

1.4 15E12 Updated Generic Troubleshooting KS

1.5 15F17 Updated Wiring Diagram with “mini” KS