TQ Analyst Gamal A. Hamid
Transcript
Page 1: Tqa

TQ AnalystGamal A. Hamid

Page 2: Tqa

Gamal A. Hamid

2

Contents

• Introduction

• Software contents

o Menu Bar

o Tool Bar

o Parameter Tabs

o Action & Option buttons

• Getting started

Page 3: Tqa

Gamal A. Hamid

3

Introduction

Page 4: Tqa

Gamal A. Hamid

4

TQA Software

• TQ Analyst is a standard Windows application .

• The software offers a complete selection of

qualitative and quantitative analytical techniques.

• It contains all of the algorithms that are typically used

for calculating component concentrations and

classifying spectra based on a set of standards.

TQA

Page 5: Tqa

Gamal A. Hamid

5

Quantitative Analysis

• Infrared spectroscopy is most often used for

qualitative identification.

• An unknown material can be determined by

comparing the infrared spectrum acquired on this

sample to the spectra of known compounds.

• Analysis of the infrared spectral band intensities as a

function of solute concentration provides a

straightforward means for determining the

concentration mixture components.

Page 6: Tqa

Gamal A. Hamid

6

Software

Page 7: Tqa

Gamal A. Hamid

7

Software Contents

The main components of software :-

I. Menu Bar

II. Tool Bar

III. Parameter Tab

IV. Action Buttons

V. Option Buttons

Page 8: Tqa

Gamal A. Hamid

8

Analyst Window

Page 9: Tqa

Gamal A. Hamid

9

I. Menu Bar

• The menus provide useful commands for working with

method files and spectral data.

• You will also find a menu of diagnostic commands that can

be used to evaluate a method as you develop it.

• Within each menu, the commands are grouped according to

their related functions

Page 10: Tqa

Gamal A. Hamid

10

1. File Menu

• Displays the commands used to create,

open, save, and print methods from TQ

Analyst and to save a TQ Analyst spectrum

in a file on your disk.

• It also explains how to open and run a

completed TQ Analyst method from other

applications and to move method files from

one computer to another.

Page 11: Tqa

Gamal A. Hamid

11

2. Edit Menu

• Default Method Path to specify a default

directory path for opening and saving.

• Default Spectrum Path to specify a default

directory path for opening and saving spectra

• Options You can customize the way TQ Analyst

displays certain windows and dialog boxes by

setting options.

• Edit Menu lets you create custom Menu

containing buttons for quickly initiating

commands.

Page 12: Tqa

Gamal A. Hamid

12

3. View Menu

• Display Setup to specify how spectra to be displayed

• Stacking spectra is useful when you are comparing

spectra that are significantly different.

• Hide Spectra to hide the selected spectra from view.

• Full Scale to adjust the vertical scale of the spectra .

• Common Scale to display all the spectra using the

same Y-axis

• Match Scale ,All the spectra in the window are

displayed using the Y-axis limits .

• Offset Scale to display the spectra vertically offset

from each other, Separating the spectra.

Page 13: Tqa

Gamal A. Hamid

13

View Menu

• Display Limits to specify the X-axis and Y-axis display limits.

• Automatic Full Scale automatically displays the spectra full.

• Absorbance at a frequency is defined by the equation

A = log(100/%T)

• Transmittance IR energy transmitted through the sample.

%T = (S/B)*100 where S is IR intensity

B is IR intensity without a sample in place (background).

• Retrieve Interferogram to display the interferogram for the

selected spectrum in the active spectral window.

• Reprocess to transform the interferogram data for the selected

spectra using different transformation parameter.

Page 14: Tqa

Gamal A. Hamid

14

4. Diagnostic Menu

• The Eigenanalysis to start the Eigenanalysis diagnostic

routine.

• The Principal Component Spectra to start the Principal

Component Spectra diagnostic routine.

• The Pairwise Concentration to start the Pairwise

Concentration diagnostic routine.

• The Spectrum Outlier to start the Spectrum Outlier

diagnostic routine.

• The Pure Component Spectra to display the synthetic pure

component spectrum for each component in a quantitative

method and one pathlength spectrum.

Page 15: Tqa

Gamal A. Hamid

15

Diagnostic Menu 2

• The Statistical Spectra to generate a series of synthetic

spectra that illustrate the variation present in the standards

and, for quantitative methods, the spectral regions that

correlate with changes in component concentration.

• Select Standards if you need help choosing appropriate

standards for a quantitative.

• The Cross Validation to start the Cross Validation

diagnostic routine.

• The External Validation to start the External Validation

diagnostic routine.

Page 16: Tqa

Gamal A. Hamid

16

Diagnostic Menu 3

• The Residual Spectra to quantify a spectrum using the active

method and then display the residual spectrum or spectra for the

spectrum that was quantified.

• The Multiple Quantify to quantify two or more spectra and

store the calculated results in a format that is accessible to other

applications.

• The Multiple Summary to quantify two or more spectra and

store a summary of the results in a format that is accessible to

other applications,

• The Quantify Sequence or Map calculate component

concentration values from a sequence (*.srs) or map (*.map) file.

Page 17: Tqa

Gamal A. Hamid

17

Diagnostic Menu 4

• The Principal Component Scores to start the Principal

Component Scores diagnostic routine.

• The PRESS to start the PRESS diagnostic routine.

• The Factor Loading to start the Factor Loading diagnostic

routine.

• The Leverage to start the Leverage diagnostic routine.

• The Loading Spectra to start the Loading Spectra diagnostic

routine.

Page 18: Tqa

Gamal A. Hamid

18

5. Window Menu and Help

• The Window menu show the names of the TQ

Analyst windows that are currently open and allow

you to make an open window active.

• Use this command to display the version number for

your TQ Analyst software

Page 19: Tqa

Gamal A. Hamid

19

II. Tool Bar

The toolbar allows you to perform operations with the

current method.

1. Calibrate

2. Quantify

3. Explain

4. Performance Index

Page 20: Tqa

Gamal A. Hamid

20

1. Calibrate

• To analyze a set of calibration standards

in order to calculate a method model for

predicting component concentrations in

unknown samples.

• The first time you calibrate a method, the

software calculates a performance index,

which appears next to the Close button

on the toolbar.

• If you change the method and then

recalibrate

Page 21: Tqa

Gamal A. Hamid

21

2. Quantify

• Using the calibration method to

analyze the unknown samples .

Page 22: Tqa

Gamal A. Hamid

22

3. Explain

TQ Analyst Explain Help

• Explain Help for TQ Analyst tab features

• Explain Help for TQ Analyst Diagnostics

• Explain Help for the Region Selection task window

• Explain Help for the Corrections task window

• Explain Help for the Calibration Results window

Page 23: Tqa

Gamal A. Hamid

23

4. Performance Index

• The performance index for the previous calibration as well

as the index for the new calibration.

• If your method has no validation standards, the symbol

"N/A," meaning "not available," is displayed in the toolbar

instead of a performance index.

• The indicator to the right of the performance index tells

whether the selected method is calibrated.

• If the method has never been calibrated, the bar is red and

the message reads "uncalibrated." After the first calibration,

the bar turns green and the "calibrated" message appears.

Page 24: Tqa

Gamal A. Hamid

24

III. Parameter Tabs

• Method parameters are grouped by function.

• Each group of parameters is contained on a file card which has a tab that is always visible

at the top of the method window.

• To display a group of parameters, click the corresponding tab.

Page 25: Tqa

Gamal A. Hamid

25

IV. Action Buttons

• These buttons, called "action buttons," allow you to perform tasks related to the

selected tab.

• For example, the "Suggest" button shown in the illustration below may be used to help

you set the Analysis Type parameter on the description tab.

Page 26: Tqa

Gamal A. Hamid

26

V. Option Buttons

• Some parameters, such as the Analysis Type and

Pathlength Type, provide a group of option buttons

and allow you to select only one of the available

options.

• You can select an option from a group of option

buttons by using the mouse to click the

corresponding button.

Page 27: Tqa

Gamal A. Hamid

27

1. Description Tab

• Use the parameters on this tab to enter general information

about your method.

• A method can be described by its title, developer, revision

number, revision date, and analysis type.

• If you need help choosing the proper analysis type for your

samples, use the Analysis Type wizard. To start the wizard,

click the Suggest Analysis Type button on this tab.

• The Description tab is displayed for all analysis types.

• The information on this tab is saved automatically when you

save the active method.

Page 28: Tqa

Gamal A. Hamid

28

Description Action Buttons

• the "Suggest" button used to help you set the Analysis Type parameter on the

Description tab.

• If you need help choosing the analysis type, use the Analysis Type wizard. To start the

wizard, click the Suggest Analysis Type button at the top of this tab.

• Display the How To window if you are looking for quick instructions on how to use the

tools in this window to complete a task. a help window that contains information about

creating TQ Analyst methods is displayed on the screen.

Page 29: Tqa

Gamal A. Hamid

29

Description Options Buttons

• The Analysis Type parameter defines the type of

results the method will produce when you use it to

analyze an unknown sample spectrum and the

algorithm that will be used for the analysis.

• There are three basic types of analyses you can do

with TQ Analyst:

Page 30: Tqa

Gamal A. Hamid

30

Analysis Type

1. Quantitative Analysis

• Measuring component amounts or concentrations in a sample mixture.

2. Classification

• Classifying an unknown sample spectrum by finding the spectrum or group of spectra

that most closely match the sample spectrum or by verifying that the sample spectrum is

similar to the spectra in a specified group.

3. Measurement

• Measuring peak heights, peak areas, peak locations, or peak widths in a sample

spectrum or calculating the ratio of two measured peaks.

Page 31: Tqa

Gamal A. Hamid

31

2. Pathlength Tab

• Sample pathlength is important to quantitative

analyses because the intensities of the peaks in

a sample spectrum are determined by the

thickness as well as the composition of the

sample.

• Pathlength is important to a spectral

measurement analysis only when the

measurement data are used for quantitative or

classification purposes.

Page 32: Tqa

Gamal A. Hamid

32

Pathlength Action Buttons

• Use Suggest Pathlength Type if you are not familiar with the pathlength

treatments provided in TQ Analyst or need help deciding which treatment would

work best for the samples.

• Use the Advanced button to access the temperature/pressure and

dilution/extraction features of TQ Analyst software.

• Use Edit Region when you need to specify or adjust the pathlength peak for the

Internal Reference or Peak Ratio pathlength type.

Page 33: Tqa

Gamal A. Hamid

33

Pathlength Option Buttons

• Undecided allows you to postpone choosing the

pathlength type.

• You cannot calibrate a method when the pathlength

type is set to Undecided.

• Constant type which uses one pathlength value to

calculate component amounts or concentrations in

all of the samples to be analyzed.

• Known an option for pathlength type if you are

analyzing samples that have different but known

thicknesses.

Page 34: Tqa

Gamal A. Hamid

34

3. Components Tab

• The components will be listed in a table.

• Each component is described by its name and an

abbreviated name as well as the number of digits and

the unit which will be used to report its

concentration.

• Concentration limits, such as analysis limits and

acceptance limits, may also be specified for each

component.

• The Components tab is displayed only when the

Analysis Type parameter on the Description tab is set

to a quantitative analysis type,

Page 35: Tqa

Gamal A. Hamid

35

Components Action Buttons

• This button starts the Assess Feasibility wizard to help you determine if there is

sufficient variability in your sample data that correlates with differences in sample

composition

• Use Assess Feasibility to quickly determine if it is possible to create a calibration

model that can calculate the concentrations of the components you want to measure.

Page 36: Tqa

Gamal A. Hamid

36

Components Option Buttons

• Concentration / Error limits, specify analysis

limits for each component in the method.

• Use Concentration Constraints option if you know

all the components in your sample and want their

predicted concentrations to sum to a 100%.

• Use Component Interaction to indicate whether

you expect any interactions between the components

you want to measure.

• Composite Components is an additional method

component that is computed from the results for one

or more measured components.

Page 37: Tqa

Gamal A. Hamid

37

4. Standards Tab

• Use the parameters on this tab to describe the standards that

will be used to calibrate, validate, or correct the active method.

• The standards will be listed in a table.

• Each standard is described by its title and usage.

• Some analysis types require additional information for each

standard, such as the concentrations or amounts of the

components the standard contains or the class the standard

belongs to.

• The Standards tab is always displayed unless the Analysis Type

parameter on the Description tab is set to Measurement Only.

Page 38: Tqa

Gamal A. Hamid

38

Standards Action Buttons

• Suggest Standards wizard to help you select the number and component concentrations of

the standards needed to calibrate and validate a quantitative method.

• Evaluate Standards wizard to determine whether the number and component conc. of the

standards are appropriate for the components that are specified in the method.

• Open Standard to select a spectral data file to use as a standard in the active method.

• View Standards to display the spectra of selected standards in your method.

• Sort Standards to rearrange the entries in the Standards table.

• Ignore Missing Data to reassign to Ignore the usage of any method standards that contain

the PLS missing data indicator value.

Page 39: Tqa

Gamal A. Hamid

39

Standards Option Buttons

• Show Spectrum Title ,File Names to specify how the standards will be listed in the

Standards table and in other tables within the TQ Analyst software. You may list the

standards by their file names, their titles, or both.

• Restrict Y-axis Range to restrict the analysis of your sample spectra to a specified Y-

axis range.

Page 40: Tqa

Gamal A. Hamid

40

5. Spectra Tab

• Format for spectral data and to apply processing operations to the

data, such as subtraction, smoothing, and baseline correction.

• The corrections may be applied to the spectra of the method

standards as well as to any sample spectra you use the method to

analyze.

• If you want to specify more than one processing step, we recommend

setting them up in this order:

o Spectral Subtraction,

o Data Format,

o Smoothing,

o Multipoint Baseline Correction.

Page 41: Tqa

Gamal A. Hamid

41

Spectra Action Buttons

• View Standards button on this tab to display the selected standards before and after

the data have been processed.

• View Standards button on the Spectra tab to see the effects of a single processing

sequence, such as smoothing, derivatives, or multi-point baseline correction, on the

selected standards.

Page 42: Tqa

Gamal A. Hamid

42

Spectra Option Buttons

• Data Format to select a format for the

spectral data you want to analyze. “first or

second derivative spectra”.

• Spectral Subtraction allows to set up

automatic subtraction for the spectra used

with this method. Automatic subtraction

subtracts a reference spectrum from the

spectrum of each unknown sample.

• Multipoint Baseline Correction to turn on

Multipoint baseline correction and specify

what kind of correction will be used.

Page 43: Tqa

Gamal A. Hamid

43

6. Regions Tab

• The parameters on this tab to identify the spectral peaks or regions

• Each region is described by its index number, measurement type, measurement location,

baseline measurement type, and the locations of the baseline points.

• If you are creating a quantitative method, choosing regions that are appropriate for the

selected analysis type and that correlate best with component concentrations.

• Once the regions are selected, several iterations of calibrating the method and refining the

region limits, the measurement type, or the baseline points may be required in order to

achieve accurate quantitative results.

Page 44: Tqa

Gamal A. Hamid

44

Regions Action Buttons

• The Suggest Regions to starts the Regions wizard to help you select the spectral

region or regions that correlate best with the component.

• The Edit Regions to choose the proper region limits and baseline points when you

can see the spectra of the standards.

• using the Region Selection task window to specify regions rather than entering region

limits and baseline points directly into the Regions table.

Page 45: Tqa

Gamal A. Hamid

45

7. Corrections Tab

• Use Corrections to define a mathematical equation which will attempt to correct a

method's calibrated model for nonlinearity or bias.

• Displayed only with quantitative method type (ACLS, CLS, SMLR, PLS, and PCR.

• There are two types of corrections you may specify:

o Linearity corrections

o Bias corrections.

• The specified correction will be applied to the values calculated by the calibrated

method to bring them closer to the actual values.

• Each correction is described by its index number, component name, and the correction

coefficients for each term in the equation for the correction curve.

Page 46: Tqa

Gamal A. Hamid

46

Corrections Action Buttons

• The Suggest Corrections starts the Corrections wizard to help you determine

whether corrections are needed for the components in your method and, if so, the

type and degree of correction you should use.

• The Edit Corrections to display the Corrections task window for editing the

correction parameters for one or more components in your method.

• The Cross Corrections to access the features for setting up a cross correction in

your method.

Page 47: Tqa

Gamal A. Hamid

47

Corrections Option Buttons

• The Method Correction to specify whether the

active method will include a post-prediction

correction.

• The Correction Standards to specify whether all

of your component correction curves will be

generated from a separate set of correction

standards or from a combination of correction

standards and calibration standards.

Page 48: Tqa

Gamal A. Hamid

48

8. Report Tab

• Use the parameters on this tab to select the

information you want to include in the sample

reports and to turn TQ Analyst’s sample checking

features on and off.

• The Report tab is displayed for all method types.

• The information on this tab is saved automatically

when you save the active method.

Page 49: Tqa

Gamal A. Hamid

49

Report Option Buttons

Page 50: Tqa

Gamal A. Hamid

50

Creating a Method

Page 51: Tqa

Gamal A. Hamid

51

Basic Types of Analyses

Quantitative

• Measuring component concentrations in a sample mixture.

Classification

• Classifying an unknown sample spectrum by finding the

spectrum or group of spectra that most closely match the

sample spectrum.

Measurement

• Measuring peak heights or areas, peak locations, or peak

widths in a sample spectrum or calculating the ratio of two

measured peaks.

Page 52: Tqa

Gamal A. Hamid

52

Creating a Quantitative Analysis

• The File menu to create a new method window.

• The Save Method As command in the File menu

to create a method file.

• The Description tab and enter a title for your

method. Choose an option for the quantitative

analysis options on the Description tab.

Page 53: Tqa

Gamal A. Hamid

53

1. Quantitative Analysis

• Select the Simple Beer’s Law analysis type when

each of the components you want to measure

produces a unique peak in the spectrum of the

sample mixture.

• Select the Classic least Squares CLS analysis type

when each component you want to measure

produces a peak or combination of peaks in the

spectrum of the sample mixture, but the component

peaks overlap significantly.

Page 54: Tqa

Gamal A. Hamid

54

2. Pathlength

• The Pathlength tab and select an option for the

Pathlength parameter.

• The Advanced button displays a window that

allows you to access the temperature/pressure and

dilution/extraction features of TQ Analyst.

• Select Constant pathlength type if you are using a

sampling accessory that has a fixed pathlength, such

as a liquid transmission cell.

Page 55: Tqa

Gamal A. Hamid

55

3. Components

• the Components tab and specify the

components you want to measure and the

analysis limits for each component.

Page 56: Tqa

Gamal A. Hamid

56

4. Standards

• Describe the standards that will be used to calibrate, validate.

Page 57: Tqa

Gamal A. Hamid

57

View Standards

Page 58: Tqa

Gamal A. Hamid

58

5. Spectra

• Format for spectral data and to apply processing operations to the data, such as

subtraction, smoothing, and baseline correction.

• The corrections may be applied to the spectra of the method standards as well as

to any sample spectra you use the method to analyze.

• If you want to specify more than one processing step, we recommend setting

them up in this order:

o Spectral Subtraction,

o Data Format,

o Smoothing,

o Multipoint Baseline Correction.

Page 59: Tqa

Gamal A. Hamid

59

6. Regions

• The Regions tab to see the spectral peaks or regions

Page 60: Tqa

Gamal A. Hamid

60

7. Calibrate

• Click the Calibrate button on the toolbar to calibrate your method.

Page 61: Tqa

Gamal A. Hamid

61

8. Quantify

Page 62: Tqa

Gamal A. Hamid

62

9. Report

• Click the Report tab and specify the

information you want to include in

your sample reports, including any

spectrum or result warnings.

Page 63: Tqa

Gamal A. Hamid

63

10. Save the method

• The Save Method command

in the File menu to save your

completed method.

Page 64: Tqa

Thanks Gamal A. Hamid