+ All Categories
Home > Documents > bmd Crossvent 3

bmd Crossvent 3

Date post: 31-Dec-2015
Category:
Upload: brett-rose
View: 42 times
Download: 1 times
Share this document with a friend
Description:
bmd Crossvent 3. Modes – Touch Screen. Adult to pediatric range Assist/Control – Volume Control Assist/Control – Pressure Control Sighs SIMV Pressure Support CPAP. Turn knobs for setting:. Maximum pressure PEEP Flow. LCD Touch Screen Select Mode then press MAIN menu. - PowerPoint PPT Presentation
21
1 bmd Crossvent 3 bmd Crossvent 3
Transcript
Page 1: bmd Crossvent 3

11

bmd Crossvent 3bmd Crossvent 3

Page 2: bmd Crossvent 3

22

Modes – Touch ScreenModes – Touch Screen

Adult to pediatric rangeAdult to pediatric range Assist/Control – Volume ControlAssist/Control – Volume Control Assist/Control – Pressure ControlAssist/Control – Pressure Control SighsSighs SIMVSIMV Pressure SupportPressure Support CPAPCPAP

Page 3: bmd Crossvent 3

33

Turn knobs for setting:Turn knobs for setting:

Maximum pressureMaximum pressure PEEPPEEP FlowFlow

Page 4: bmd Crossvent 3

44

LCD Touch ScreenLCD Touch ScreenSelect Mode then press MAIN menuSelect Mode then press MAIN menu Then press Then press

Rate & TV Rate & TV – adjust with adjust with

up/down up/down arrowsarrows

IT, ET & I:E IT, ET & I:E Ratio Ratio displayeddisplayed

Flow, PEEP Flow, PEEP & Max & Max Press limitPress limit– adjust with adjust with

dials dials

Page 5: bmd Crossvent 3

55

On the left sideOn the left side

ON/OFFExternal Power Supply

Alarm Reset

LED ChargingLight

Pneumotach/FlowSensor Input

Exhalation ValveConnector

Airway PressureConnector

Pt Circuit

O2 Sensor

Page 6: bmd Crossvent 3

66

bmd CrossVent 3bmd CrossVent 3

powered by a rechargeable NiMH powered by a rechargeable NiMH batterybattery– lasts 6 hourslasts 6 hours

BioMed 16vDC 3A charger/converter BioMed 16vDC 3A charger/converter recharges internal battery and allows recharges internal battery and allows for AC operationfor AC operation

LCD screen has LCD charging LCD screen has LCD charging indicator and visually large LOW indicator and visually large LOW BATTERY alarmBATTERY alarm

Page 7: bmd Crossvent 3

77

Flow Sensor at proximal Flow Sensor at proximal airwayairway

Page 8: bmd Crossvent 3

88

Flow SensorFlow Sensor

Page 9: bmd Crossvent 3

99

On the right sideOn the right side

Gas Supply Inlet (31 - 75 psig or 45 - 75 when using blender)

Gas EntrainmentON/OFF

Bleed Exhaust

Page 10: bmd Crossvent 3

1010

Gas EntrainmentGas Entrainment

When connected to 100% oxygen When connected to 100% oxygen source gas with the Gas Entrainment source gas with the Gas Entrainment ON, the oxygen concentration is ON, the oxygen concentration is diluted to 50%.diluted to 50%.

There will be some decrease in There will be some decrease in delivered flow rate with Entrainment delivered flow rate with Entrainment ON under high load conditionsON under high load conditions

Page 11: bmd Crossvent 3

1111

The rear panelThe rear panel

Maximum pressure relief valve (120 Maximum pressure relief valve (120 cm Hcm H22O)O)

Negative pressure relief valve – Negative pressure relief valve – opens at - 4 cm H2O with machine opens at - 4 cm H2O with machine shutdownshutdown

Page 12: bmd Crossvent 3

1212

SIMV settingsSIMV settings

Page 13: bmd Crossvent 3

1313

Press arrows to increase or Press arrows to increase or decrease Rate or Tidal Volumedecrease Rate or Tidal Volume

Page 14: bmd Crossvent 3

1414

Touch screen to highlight TV – Touch screen to highlight TV – Adjust with up/down arrowsAdjust with up/down arrows

Page 15: bmd Crossvent 3

1515

CPAP ModeCPAP Mode

Page 16: bmd Crossvent 3

1616

ALARMSALARMS

ALARM 1 ALARM 1 – Peak PressurePeak Pressure– Respiratory RateRespiratory Rate– Exhaled Tidal Volume (can be switched off)Exhaled Tidal Volume (can be switched off)– Exhaled Minute Volume (can be switched Exhaled Minute Volume (can be switched

off)off) ALARM 2ALARM 2

– PEEPPEEP– Mean Pressure (can be switched off)Mean Pressure (can be switched off)– Oxygen (can be switched off)Oxygen (can be switched off)

Page 17: bmd Crossvent 3

1717

To set Alarms Press ALARM To set Alarms Press ALARM 11

Press Press High or High or Low Low Limits to Limits to ChangeChange

Adjust Adjust with with up/down up/down arrowsarrows

Page 18: bmd Crossvent 3

1818

Repeat for ALARM 2 settingsRepeat for ALARM 2 settings

Page 19: bmd Crossvent 3

1919

33rdrd Level Alarms Level Alarms

Low Gas Supply Pressure

Low Battery

Ventilator Failure

Page 20: bmd Crossvent 3

2020

With Alarm Condition:With Alarm Condition:

the value for the exceeded limitflashes red

-1

Page 21: bmd Crossvent 3

2121

When power off, press and When power off, press and hold ALARM RESEThold ALARM RESET


Recommended